Mobile C-arm units, so named because of their configuration, are used primarily for fluoroscopic imaging during surgical, orthopedic, critical care, and emergency care procedures. They also provide radiographic capability, although they are rarely used for this imaging modality. We evaluated three mobile C-arm units from three manufacturers. We based our ratings on the ability of the units to safely and reliably produce the best possible image quality while delivering the lowest possible radiation dose to the patient and protecting personnel from radiation exposure; we also considered human factors design. The Philips BV 25 N is rated Acceptable and is preferred as a general-purpose unit. The OEC-Diasonics 9000/6 is also rated Acceptable and may be preferred when the hospital has specific uses for its optional specialty modules. The Fischer Ortho-Omni 325 is rated Acceptable-Not Recommended because of its relatively high patient exposure rates, poorer object thickness compensation (penetration) capability, poorer spatial resolution, and overall poor human factors design compared with the other evaluated units. Also see "Issues of Radiation Exposure" and "Ensuring Optimum Performance."
移动C形臂设备,因其配置而得名,主要用于手术、骨科、重症监护和急救程序中的透视成像。它们也具备放射成像能力,尽管很少用于这种成像方式。我们评估了来自三家制造商的三款移动C形臂设备。我们的评级基于这些设备在为患者提供尽可能低的辐射剂量并保护人员免受辐射暴露的同时,安全可靠地产生最佳图像质量的能力;我们还考虑了人为因素设计。飞利浦BV 25 N被评为合格,作为通用设备更受青睐。OEC - Diasonics 9000/6也被评为合格,当医院对其可选的专业模块有特定用途时可能更受青睐。菲舍尔Ortho - Omni 325被评为合格但不推荐,因为与其他评估的设备相比,其患者照射率相对较高,物体厚度补偿(穿透)能力较差,空间分辨率较差,且整体人为因素设计不佳。另见“辐射暴露问题”和“确保最佳性能”。