骨整合式植入物系统在上肢截肢患者中的稳定固定:55 例 RSA 与放射学评估中的迁移与骨重塑。

Stable fixation of an osseointegated implant system for above-the-knee amputees: titel RSA and radiographic evaluation of migration and bone remodeling in 55 cases.

机构信息

Harris Orthopaedic Laboratory, Massachusetts General Hospital, Boston, USA.

出版信息

Acta Orthop. 2012 Apr;83(2):121-8. doi: 10.3109/17453674.2012.678799. Epub 2012 Apr 11.

Abstract

BACKGROUND AND PURPOSE

Rehabilitation of patients with transfemoral amputations is particularly difficult due to problems in using standard socket prostheses. We wanted to assess long-term fixation of the osseointegrated implant system (OPRA) using radiostereometric analysis (RSA) and periprosthetic bone remodeling.

METHODS

51 patients with transfemoral amputations (55 implants) were enrolled in an RSA study. RSA and plain radiographs were scheduled at 6 months and at 1, 2, 5, 7, and 10 years after surgery. RSA films were analyzed using UmRSA software. Plain radiographs were graded for bone resorption, cancellization, cortical thinning, and trabecular streaming or buttressing in specifically defined zones around the implant.

RESULTS

At 5 years, the median (SE) migration of the implant was -0.02 (0.06) mm distally. The rotational movement was 0.42 (0.32) degrees around the longitudinal axis. There was no statistically significant difference in median rotation or migration at any follow-up time. Cancellization of the cortex (plain radiographic grading) appeared in at least 1 zone in over half of the patients at 2 years. However, the prevalence of cancellization had decreased by the 5-year follow-up.

INTERPRETATION

The RSA analysis for the OPRA system indicated stable fixation of the implant. The periprosthetic bone remodeling showed similarities with changes seen around uncemented hip stems. The OPRA system is a new and promising approach for addressing the challenges faced by patients with transfemoral amputations.

摘要

背景与目的

由于标准接受腔假肢存在问题,股骨干截肢患者的康复尤其困难。我们希望通过放射学体层摄影术(RSA)和假体周围骨重塑评估骨整合植入物系统(OPRA)的长期固定情况。

方法

我们对 51 名股骨干截肢患者(55 个植入物)进行了 RSA 研究。RSA 和普通 X 光片在手术后 6 个月、1、2、5、7 和 10 年进行。使用 UmRSA 软件分析 RSA 胶片。对普通 X 光片进行骨吸收、骨溶解、皮质变薄以及植入物周围特定区域的小梁流或支撑的分级。

结果

在 5 年时,植入物的中位数(SE)迁移距离为-0.02(0.06)mm 远侧。旋转运动为围绕纵轴的 0.42(0.32)度。在任何随访时间,旋转或迁移的中位数均无统计学差异。至少有一半患者在 2 年内至少有 1 个区域出现皮质骨溶解(普通 X 线分级)。然而,到 5 年随访时,皮质骨溶解的发生率已降低。

解释

OPRA 系统的 RSA 分析表明植入物的固定稳定。假体周围骨重塑与未固定髋关节假体周围的变化相似。OPRA 系统是一种解决股骨干截肢患者所面临挑战的新的、有前途的方法。