Suppr超能文献

使用活检钳成功取出股骨动脉中断裂并嵌顿的 0.035 英寸 Terumo 线。

Successful retrieval of a fractured and entrapped 0.035-inch terumo wire in the femoral artery using biopsy forceps.

机构信息

Cardiovascular Center, Korea University Guro Hospital, Seoul, Korea.

出版信息

Korean Circ J. 2012 Mar;42(3):201-4. doi: 10.4070/kcj.2012.42.3.201. Epub 2012 Mar 26.

Abstract

A 0.035-inch guide wire fracture and entrapment in a peripheral artery is a very rare complication, but when it does occur it may lead to life-threatening complications, such as perforation, thrombus formation, embolization, and subsequent limb ischemia. We describe our experience of successfully retrieving a fractured 0.035-inch Terumo guide wire in the external iliac artery using a biopsy forcep.

摘要

0.035 英寸导丝断裂并嵌顿在外周动脉是一种非常罕见的并发症,但一旦发生,可能导致危及生命的并发症,如穿孔、血栓形成、栓塞和随后的肢体缺血。我们描述了使用活检钳成功取出断裂的 0.035 英寸 Terumo 导丝的经验。