Suppr超能文献

Determination of Total Selenium in Infant Formulas: Comparison of the Performance of FIA and MCFA Flow Systems.

作者信息

Pistón Mariela, Knochen Moisés

机构信息

Department of Analytical Chemistry, School of Chemistry, Universidad de la República (UdelaR), Avenue Gral. Flores 2124, 11800 Montevideo, Uruguay.

出版信息

Int J Anal Chem. 2012;2012:918292. doi: 10.1155/2012/918292. Epub 2012 Feb 20.

Abstract

Two flow methods, based, respectively, on flow-injection analysis (FIA) and on multicommutated flow analysis (MCFA), were compared with regard to their use for the determination of total selenium in infant formulas by hydride-generation atomic absorption spectrometry. The method based on multicommutation provided lower detection and quantification limits (0.08 and 0.27 μg L(-1) compared to 0.59 and 1.95 μ L(-1), resp.), higher sampling frequency (160 versus. 70 samples per hour), and reduced reagent consumption. Linearity, precision, and accuracy were similar for the two methods compared. It was concluded that, while both methods proved to be appropriate for the purpose, the MCFA-based method exhibited a better performance.

摘要
https://cdn.ncbi.nlm.nih.gov/pmc/blobs/0f8a/3296178/d45524601fa6/IJAC2012-918292.001.jpg