Graduate Program in Chemistry, Facultad de Química, Universidad de la República, Avda. Gral. Flores 2124, Montevideo 11200, Uruguay.
Grupo de Análisis de Elementos Traza y Desarrollo de Estrategias Simples para Preparación de Muestras (GATPREM), Química Analítica (DEC), Facultad de Química, Universidad de la República, Montevideo 11200, Uruguay.
Molecules. 2021 Mar 16;26(6):1662. doi: 10.3390/molecules26061662.
An on-line preconcentration system for the simultaneous determination of Copper (Cu) and manganese (Mn) in water samples was developed and coupled to a microwave-induced plasma optical emission spectrometer (MIP OES). The flow injection system was designed with a minicolumn packed with sisal fiber (). A multivariate experimental design was performed to evaluate the influence of pH, preconcentration time, and eluent concentration. Optimal conditions for sample preparation were pH 5.5, preconcentration time was 90 s, and HCl 0.5 mol L was the eluent. The main figures of merit were detection limits 3.7 and 9.0 µg L for Cu and Mn, respectively. Precision was expressed as a relative standard deviation better than 10%. Accuracy was evaluated via spiked recovery assays with recoveries between 75-125%. The enrichment factor was 30 for both analytes. These results were adequate for water samples analysis for monitoring purposes. The preconcentration system was coupled and synchronized with the MIP OES nebulizer to allow simultaneous determination of Cu and Mn as a novel sample introduction strategy. The sampling rate was 20 samples/h. Sisal fiber resulted an economical biosorbent for trace element preconcentration without extra derivatization steps and with an awfully time of use without replacement complying with the principles of green analytical methods.
建立了一种在线预浓缩系统,用于同时测定水样中的铜(Cu)和锰(Mn),并将其与微波诱导等离子体光学发射光谱仪(MIP OES)耦合。该流动注射系统设计有一个用剑麻纤维填充的微型柱()。采用多元实验设计来评估 pH 值、预浓缩时间和洗脱液浓度的影响。样品制备的最佳条件为 pH 5.5、预浓缩时间 90 s 和 HCl 0.5 mol L 作为洗脱液。主要的衡量标准是 Cu 和 Mn 的检测限分别为 3.7 和 9.0 µg L。精密度表示为相对标准偏差优于 10%。通过加标回收试验评估准确性,回收率在 75-125%之间。两种分析物的富集因子均为 30。这些结果足以用于监测目的的水样分析。该预浓缩系统与 MIP OES 雾化器耦合和同步,允许作为一种新的样品引入策略同时测定 Cu 和 Mn。采样率为 20 个样品/小时。剑麻纤维是一种经济的生物吸附剂,用于痕量元素的预浓缩,无需额外的衍生化步骤,使用时间长,无需更换,符合绿色分析方法的原则。