Zhang Yanlin, Adeloju Samuel B
NanoScience and Sensor Technology Research Group, School of Applied Sciences and Engineering, Monash University Gippsland Campus, Churchill, Victoria 3842, Australia.
Talanta. 2008 Aug 15;76(4):724-30. doi: 10.1016/j.talanta.2008.03.056. Epub 2008 Apr 16.
A simple and robust flow injection system which permits low sample and reagent consumption is described for rapid and reliable hydride generation atomic absorption spectrometric determination of selenium, arsenic and bismuth. The system, which composed of one peristaltic pump and one four channel solenoid valve, used water as the carrier streams for both sample and NaBH(4) solution. Rapid off-line pre-reduction of the analytes was achieved by using hydroxylamine hydrochloride for selenium and a mixture of potassium iodide and ascorbic acid for arsenic and bismuth. Transition metal interference was eliminated with the addition of thiourea and EDTA into the NaBH(4) solution and significant sensitivity enhancement was observed for selenium in the presence of thiourea in the reductant solution. Under optimised conditions, the method achieved detection limits of 0.2 ng mL(-1) for Se, 0.5 ng mL(-1) for As and 0.3 ng mL(-1) for Bi. The method was very reproducible, achieving relative standard deviations of 6.3% for Se, 3.6% for As and 4.7% for Bi, and has a sample throughput of 360 h(-1). Successful application of the method to the quantification of selenium, arsenic and bismuth in a certified reference river sediment sample is reported.
本文描述了一种简单且耐用的流动注射系统,该系统可实现低样品和试剂消耗,用于快速、可靠地通过氢化物发生原子吸收光谱法测定硒、砷和铋。该系统由一个蠕动泵和一个四通道电磁阀组成,使用水作为样品和硼氢化钠(NaBH₄)溶液的载流。通过使用盐酸羟胺对硒进行预还原,以及使用碘化钾和抗坏血酸的混合物对砷和铋进行预还原,实现了分析物的快速离线预还原。通过向硼氢化钠溶液中添加硫脲和乙二胺四乙酸(EDTA)消除了过渡金属干扰,并且在还原剂溶液中存在硫脲的情况下,观察到硒的灵敏度显著提高。在优化条件下,该方法对硒的检测限为0.2 ng mL⁻¹,对砷为0.5 ng mL⁻¹,对铋为0.3 ng mL⁻¹。该方法具有很好的重现性,硒的相对标准偏差为6.3%,砷为3.6%,铋为4.7%,样品通量为360 h⁻¹。本文报道了该方法在认证的河流沉积物标准参考样品中硒、砷和铋定量分析中的成功应用。