Suppr超能文献

Venturi grafts for hemodialysis access.

作者信息

Drasler W J, Wilson G J, Jenson M L, Klement P, George S A, Protonotarios E I, Dutcher R G, Possis Z C

机构信息

Possis Medical, Inc. Minneapolis, Minnesota 55427.

出版信息

ASAIO Trans. 1990 Jul-Sep;36(3):M753-7.

PMID:2252803
Abstract

Vascular access grafts can produce venous and puncture site stenosis, and excessive shunted flow. A vascular access graft that controls the shunted flow at 500-650 ml/min by venturi flow resistance, and offers the potential for reduced venous stenosis, is presented. Twelve venturi grafts consisting of 6 mm ePTFE, with a thermally formed venturi resistance, were implanted in a canine femoral arterio-venous (A-V) model and cannulated for simulated dialysis for periods up to 6 months. Pressures, bleeding times, and puncture site healing were compared at sites upstream and downstream of the venturi. Nine 6 mm ePTFE grafts were implanted as controls. Venturi and control graft patency rates were similar: 90% for venturi grafts and 89% for control implants at 4 months. Measurements verified pressure dissipation by the venturi from 85 to 15 mmHg, delivering near-normal pressures to the vein; downstream pressures were 22 mmHg higher in controls. Sites downstream of the venturi did not bleed at 4 min, whereas bleeding was often noted after 9 min at other sites. Puncture sites upstream of the venturi had less stenosis and less mural thrombus than sites downstream of the venturi, and compared favorably with puncture sites in the control grafts. This improved healing and low pressure delivery to the vein may offer clinical advantages.

摘要

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验