Suppr超能文献

一种作为可穿刺动静脉分流器的胶原涂层织物血管假体。

A collagen coated fabric vascular prosthesis as a punctureable A-V shunt.

作者信息

Noishiki Y, Marat D, Yamane Y, Satoh S, Ma X H, Iwai Y, Mo M, Matsumoto A

机构信息

First Department of Surgery, Yokohama City University School of Medicine, Japan.

出版信息

ASAIO J. 1996 Sep-Oct;42(5):M687-93. doi: 10.1097/00002480-199609000-00075.

Abstract

A fabric vascular prosthesis sealed with succinylated collagen (SC) was developed as an arteriovenous (A-V) shunt graft for hemodialysis. The SC graft was soft, pliable, flexible, and puncturable, with quick hemostasis. A needle puncture made a smaller hole in the SC graft wall than in a control expanded-polytetrafluoroethylene (e-PTFE) graft such as is usually used for an A-V shunt. The SC graft was extremely hydrous, because it is sealed with water, which is absorbed into the intermolecular spaces of negatively charged collagen. The SC suspension was injected with pressure into a knitted fabric vascular prosthesis wall (water permeability, 1,200 ml) so as to become entangled in the Dacron network. The graft then was lyophilized and thermally cross-linked. Water leakage from six holes created by an 18 G needle puncture in vitro under water pressure of 120 mmHg was 34.5 +/- 29.9 ml/min in the SC graft and 169.9 +/- 38.5 ml/min in the control e-PTFE graft. Hemostatic time at six 18G needle puncture sites on grafts implanted in the abdomen of 12 dogs was 4.5 +/- 2.5 mins in the SC graft and 34.2 +/- 11.5 mins in the control graft. After implantation, the luminal surface of the SC grafts had a thinner thrombus layer than did the control grafts. After 1 week, a thin thrombus layer covered the luminal surface of the SC grafts, and puncture sites were recognized as small dots under it, but in the control grafts, the thrombus layer was thick and the puncture sites could not be seen. These results suggest that the SC graft is less thrombogenic and has less blood leakage from the puncture sites than does the control graft.

摘要

一种用琥珀酰化胶原蛋白(SC)密封的织物血管假体被开发用作血液透析的动静脉(A-V)分流移植物。SC移植物柔软、柔韧、可弯曲且可穿刺,止血迅速。与通常用于A-V分流的对照膨胀聚四氟乙烯(e-PTFE)移植物相比,针刺在SC移植物壁上形成的孔更小。SC移植物含水量极高,因为它是用水密封的,水被吸收到带负电荷的胶原蛋白的分子间隙中。将SC悬浮液在压力下注入针织织物血管假体壁(水渗透性为1200毫升),使其缠结在涤纶网中。然后将移植物冻干并进行热交联。在体外120 mmHg水压下,18G针刺造成的六个孔的漏水情况在SC移植物中为34.5±29.9毫升/分钟,在对照e-PTFE移植物中为169.9±38.5毫升/分钟。在12只狗腹部植入的移植物上,六个18G针刺部位的止血时间在SC移植物中为4.5±2.5分钟,在对照移植物中为34.2±11.5分钟。植入后,SC移植物的管腔表面血栓层比对照移植物薄。1周后,一层薄血栓层覆盖了SC移植物的管腔表面,穿刺部位在其下方可被识别为小点,但在对照移植物中,血栓层很厚,穿刺部位无法看到。这些结果表明,与对照移植物相比,SC移植物的血栓形成性较低,穿刺部位的血液渗漏较少。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验