Department of Biomedical Engineering, Chung Yuan Christian University, Chung-Li, Taiwan.
J Hum Hypertens. 2013 Mar;27(3):204-10. doi: 10.1038/jhh.2012.17. Epub 2012 May 3.
Central aortic systolic blood pressure (SBP-C) can be estimated from a cuff oscillometric waveform derived during the pulse volume plethysmography (PVP) by applying a device-specific aortic pressure-to-PVP waveform-generalized transfer function (A2P(GTF)). The present study compared the performance of an aortic-to-brachial pressure waveforms generalized transfer function (A2B(GTF)), which is independent of any PVP devices, with an A2P(GTF). Generalized transfer function of aortic-to-brachial (A2B(GTF)) and aortic-to-PVP (A2P(GTF)) were generated from the simultaneously obtained central aortic and brachial pressure waveforms recorded by a high-fidelity dual pressure sensor catheter, and the PVP waveform recorded by a customized noninvasive blood pressure monitor during cardiac catheterization in 40 patients, and were then applied in another 100 patients with simultaneously recorded invasive aortic pressure and noninvasively calibrated (using cuff SBP and diastolic blood pressures) PVP waveforms. The mean difference±s.d. between the noninvasively estimated and invasively recorded SBP-C was -2.1±7.7 mm Hg for A2B(GTF), which was not greater than that of -3.0±7.7 mm Hg for A2P(GTF) (P<0.01). In conclusion, SBP-C can be measured reliably using a noninvasive blood pressure monitor by applying either an A2P(GTF) or A2B(GTF) to a noninvasively calibrated PVP waveform. The performance of an A2B(GTF) is not inferior to that of an A2P(GTF).
中心主动脉收缩压(SBP-C)可通过应用特定设备的主动脉压力-脉搏体积描记图(PVP)波形通用传递函数(A2P(GTF))从 PVP 衍生的袖带示波法测量的主动脉收缩压波形中估计。本研究比较了一种不依赖任何 PVP 设备的主动脉-肱动脉压力波形通用传递函数(A2B(GTF))与 A2P(GTF)的性能。通过同时获得的高保真度双压力传感器导管记录的中心主动脉和肱动脉压力波形以及定制的无创血压监测仪记录的 PVP 波形生成了主动脉-肱动脉(A2B(GTF))和主动脉-脉搏体积描记图(A2P(GTF))的通用传递函数,然后将其应用于另外 100 例同时记录有创主动脉压力和经无创校准(使用袖带 SBP 和舒张压)的 PVP 波形的患者。应用 A2B(GTF)和 A2P(GTF)时,无创估计的 SBP-C 与有创记录的 SBP-C 之间的平均差值±标准差分别为-2.1±7.7 mmHg 和-3.0±7.7 mmHg(P<0.01)。总之,通过将 A2P(GTF)或 A2B(GTF)应用于无创校准的 PVP 波形,使用无创血压监测仪可以可靠地测量 SBP-C。A2B(GTF)的性能并不逊于 A2P(GTF)。