Huang Jiapeng, Dodwani Kishin, Simon Paul, Zhou Jing
Department of Anesthesia, Jewish Hospital and St. Mary's Healthcare, Louisville, KY, USA.
J Anaesthesiol Clin Pharmacol. 2012 Apr;28(2):230-1. doi: 10.4103/0970-9185.94902.
We describe a case where the expiratory bacterial filter was completely blocked by a plastic membrane. Significantly increased airway pressures and inability to deliver adequate tidal volume to the patient were experienced. The cause of machine failure could not be identified until a complete preanesthesia machine manual check was performed. All machines and circuits should be checked manually, even with the availability of new and advanced self-tests in machines. Backup ventilation equipments and familiarity with signs of expiratory and inspiratory limb obstruction should not be underestimated.