State Key Laboratory of Ultrasound Engineering in Medicine, Department of Biomedical Engineering, Chongqing Medical University, Chongqing 400016, China.
Ultrasonics. 2013 Jan;53(1):77-83. doi: 10.1016/j.ultras.2012.04.001. Epub 2012 Apr 27.
The purpose of this study was to evaluate pathological changes of the portal vein (PV) and the effects on main branches of the hepatic PV during HIFU (high-intensity focused ultrasound) sonication when liver tissue adjacent to the main branches of hepatic PV was ablated. Normal liver tissue at 0mm, 5mm, 10mm away from the hepatic portal vein in 50 healthy goats was ablated with magnetic resonance image-guided HIFU (MRgHIFU). MRI showed a non-perfusion region at the target area but did not show any significant changes of the PV immediately after HIFU. The histological examination 1 day after HIFU showed coagulative necrosis at the target area, revealed deep-dyed swelling collagen (CS) fibers and vessel wall fracture (VWF) in the PV adjacent to the target area; however, no CS or VWF was observed in the PV 1 week after HIFU ablation. The energy required to ablate the foci at 0mm was 21% more than that at 10mm from the PV (p<0.05); the energy needed to ablate foci 5mm away from the PV was 10% more than that at 10mm from the PV (p<0.05). We concluded that minor injury of the hepatic portal vein may occur when ablating the adjacent liver tissue, and the acoustic energy deposition is related to the distance to the portal vein.
本研究旨在评估当肝门静脉(PV)主要分支附近的肝组织被消融时,高强度聚焦超声(HIFU)声处理对门静脉的病理变化及其对肝门静脉主要分支的影响。采用磁共振图像引导 HIFU(MRgHIFU)消融 50 只健康山羊距离肝门静脉 0mm、5mm 和 10mm 处的正常肝组织。MRI 显示在靶区出现无灌注区域,但 HIFU 后即刻 PV 未显示任何明显变化。HIFU 后 1 天的组织学检查显示靶区发生凝固性坏死,PV 紧邻靶区的深染肿胀胶原(CS)纤维和血管壁破裂(VWF);然而,PV 在 HIFU 消融后 1 周内未观察到 CS 或 VWF。与距离 PV 10mm 处的消融灶相比,0mm 处的消融灶所需能量多 21%(p<0.05);距离 PV 5mm 处的消融灶所需能量比距离 PV 10mm 处的消融灶多 10%(p<0.05)。我们得出结论,当消融毗邻肝组织时,肝门静脉可能会受到轻微损伤,声能沉积与距离门静脉的远近有关。