A method has been developed to reduce the mass spectrometric ion signal suppression associated with the use of TFA as an additive in LC mobile phases. Through post-column infusion of diluted NH(4)OH solution to LC eluents, the ammonium ion introduced causes the neutral analyte-TFA ion pair to dissociate which consequently releases the protonated analyte as free ions into the gas phase (through regular electrospray ionization mechanisms). An ion signal improvement from 1.2 to 20 times for a variety of compounds had been achieved through the application of this method. The molar ratios of NH(4)OH:TFA which result in a reduction of signal suppression were determined to be between 0.5:1 and 50:1. In addition, it was shown that this NH(4)OH infusion method could reduce the level of doubly-charged species and the product ions formed via in-source collision. The use of diluted NH(4)OH solution is favorable since it is compatible with mass spectrometry analysis, and it is applicable in both positive and negative-ion generation mode.
已经开发出一种方法来减少与在 LC 流动相中使用 TFA 作为添加剂相关的质谱离子信号抑制。通过将稀释的 NH(4)OH 溶液后注入 LC 洗脱液中,引入的铵离子导致中性分析物-TFA 离子对解离,从而将质子化的分析物作为游离离子释放到气相中(通过常规电喷雾电离机制)。通过应用这种方法,已经实现了各种化合物的离子信号提高 1.2 到 20 倍。结果表明,导致信号抑制减少的 NH(4)OH:TFA 的摩尔比在 0.5:1 至 50:1 之间。此外,还表明这种 NH(4)OH 注入方法可以减少通过源内碰撞形成的双电荷物种和产物离子的水平。使用稀释的 NH(4)OH 溶液是有利的,因为它与质谱分析兼容,并且适用于正离子和负离子生成模式。