Kim B C, Lee K B, Park T S, Lee J M, Lee S H, Oh P J, Lee M K, Ahn J K
Ionizing Radiation Center, KRISS, Yuseng-Gu, Daejeon 350-340, Republic of Korea.
Appl Radiat Isot. 2012 Sep;70(9):1934-9. doi: 10.1016/j.apradiso.2012.02.020. Epub 2012 Mar 4.
The Korea Research Institute of Standards and Science (KRISS) has developed a primary system for the activity standardization of gaseous radon-222, based on the defined solid angle counting method. The size of adsorbed radon is determined by a Cyclone Storage Phosphor System, and a buffer chamber is introduced between the measurement chamber and the source for the purpose of gas purification. The measured activity of gaseous radon-222 and its associated measurement uncertainty obtained using the system are presented.
韩国标准科学研究院(KRISS)基于定义的立体角计数法,开发了一种用于气态氡-222活度标准化的初级系统。吸附氡的量由旋风存储磷光体系统测定,并且在测量室和源之间引入了一个缓冲室,用于气体净化。给出了使用该系统测得的气态氡-222的活度及其相关的测量不确定度。