Highly cross-linked vs conventional polyethylene: no differences in rim notching from micromotion on retrieved acetabular liners.

机构信息

Division of Orthopaedic Surgery, London Health Sciences Centre, London, Ontario, Canada.

出版信息

J Arthroplasty. 2012 Oct;27(9):1616-1621.e1. doi: 10.1016/j.arth.2012.03.034. Epub 2012 May 17.

Abstract

Previous literature suggested highly cross-linked acetabular liners demonstrated notching patterns that may be a point of crack origin and rim failure. We examined (1) whether notching patterns and rim cracks existed and demonstrated similar morphological properties in retrieved highly cross-linked and non-cross-linked liners and (2) whether the dimensions of these notches correlated with their duration of implantation. We retrieved a series of 14 identical liners out to an average of 2.03 years. Liners were microscopically examined and then scanned using microcomputed tomography. All liners demonstrated identical notching patterns. Microcomputed tomographic scans demonstrated no signs of crack initiation or rim failure but were able to accurately quantify the notch dimensions. The notching patterns were likely caused by liner-cup micromotion and are better characterized as creep deformation because they did not progress markedly over longer durations of implantation.

摘要

先前的文献表明,高度交联的髋臼衬垫表现出的切痕模式可能是裂纹起源和边缘失效的一个点。我们检查了(1)在取出的高度交联和非交联衬垫中是否存在切痕模式和边缘裂纹,并表现出相似的形态特征,以及(2)这些切痕的尺寸是否与其植入时间有关。我们取出了一系列 14 个相同的衬垫,平均使用时间为 2.03 年。衬垫进行了显微镜检查,然后使用微计算机断层扫描进行了扫描。所有衬垫都表现出相同的切痕模式。微计算机断层扫描显示没有裂纹起始或边缘失效的迹象,但能够准确地量化切痕尺寸。切痕模式可能是由衬垫-杯的微动引起的,因为它们在较长的植入时间内没有明显进展,所以更适合被描述为蠕变变形。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验