Suppr超能文献

植入右心室间隔的主动固定起搏器导线导致晚期左心室穿孔。

Late left ventricular perforation by active fixation pacemaker lead implanted in the right ventricular septum.

作者信息

Kondoh Haruhiko, Funatsu Toshihiro, Taniguchi Kazuhiro

机构信息

Department of Cardiovascular Surgery, Japan Labor Health and Welfare Organization, Osaka Rosai Hospital, Nagasone-cho, Sakai, Osaka, Japan.

出版信息

J Card Surg. 2012 Jul;27(4):530-1. doi: 10.1111/j.1540-8191.2012.01479.x. Epub 2012 May 22.

Abstract

We report a rare case of late left ventricular perforation caused by a migrated pacemaker lead, which had been implanted in the right ventricular septum three months prior. Preoperative computed tomography findings revealed that the tip of the migrated pacemaker lead had perforated the ventricular wall near the intraventricular groove into the anterior chest wall.

摘要

我们报告一例罕见的晚期左心室穿孔病例,由三个月前植入右心室间隔的起搏器导线移位所致。术前计算机断层扫描结果显示,移位的起搏器导线尖端已穿破室间沟附近的心室壁,进入前胸壁。