Department of Orthopedics, Chia-Yi Christian Hospital, Chiayi, Taiwan.
Kaohsiung J Med Sci. 2012 Jun;28(6):345-9. doi: 10.1016/j.kjms.2011.11.018. Epub 2012 Apr 3.
The fracture of uncemented, fully porous, coated femoral stems is a rare complication that develops after primary total hip arthroplasty, and it is even rarer after revision surgery. To the best of our knowledge, cases of only 10 broken fully coated stems have been reported. This is the first report on fracture of the VerSys fully bead-coated femoral stem, which is designed for diaphyseal fixation and for application in revision surgery or when a proximal fixation is not feasible. Between 2000 and 2008, we implanted 41 VerSys fully coated stems in 40 patients. We present four cases of femoral stem fractures at 23, 31, 40, and 86 months after surgery. The common risk factors for these four patients were inadequate support for the implant in the proximal femur due to nonunion in this area, a stem diameter ≤12 mm, and a relatively young age.
非骨水泥、全多孔、涂层股骨柄骨折是初次全髋关节置换术后一种罕见的并发症,在翻修手术后更为罕见。据我们所知,仅有 10 例完全涂层的断裂柄报告。这是首例关于 VerSys 全珠涂层股骨柄骨折的报告,该柄设计用于骨干固定,适用于翻修手术或近端固定不可行时。在 2000 年至 2008 年间,我们在 40 名患者中植入了 41 个 VerSys 全涂层柄。我们报告了手术后 23、31、40 和 86 个月发生的 4 例股骨柄骨折。这 4 名患者的共同危险因素是由于该区域的非愈合导致近端股骨中植入物的支撑不足、柄直径≤12mm 和相对年轻。