[NIOBI-X filter: reduction of patient exposure and effect on image quality].

作者信息

Rozza M, Campoleoni M, Landini A, Orsini S

机构信息

Servizio di Radiodiagnostica e Radioterapia, Istituto L. Mangiagalli, Milano.

出版信息

Radiol Med. 1990 Nov;80(5):740-4.

PMID:2267397
Abstract

Patient protection against ionizing radiation has become more and more important and the subject has been included in all the latest ICRP publications. Recently a 50 microns Niobium filter (NIOBI-X) has come out, which, installed on tungsten anode X-ray apparatuses, allows patient dose reduction by absorbing the radiation which does not contribute to the image. In order to verify the NIOBI-X actual effect on image quality and on patient dose reduction, image resolution, image contrast and patient exposure (air-Kerma) were measured using a bone-equivalent stepped wedge. The NIOBI-X, added to the total filtration of the X-ray apparatus, allows an air-Kerma reduction of about 50% from 60 to 130 kV. In the case of bone-equivalent material exposure the filter causes a slight contrast loss and does not affect, or even improves, image resolution. Before choosing the NIOBI-X filter for all the X-ray apparatuses a cost-benefit analysis should of course be made by evaluating results and cost related to other kinds of filter. With particular regard to Aluminium filters, whose prices are far lower than Niobium ones, it has to be pointed out that: 1) in order to have the same air-Kerma reduction the Aluminium filter thickness must be quite heavy (2.8 mm) which causes practical problems of use; 2) the Aluminium filter causes image quality loss especially at lower kVs.

摘要

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验