Browne Vanessa, Flewelling Michael, Wierenga Mark, Wilson Alisa, Aprecio Ray, Richardson Paul, Angelov Nikola, Johnson Neal
Center for Dental Research, Loma Linda University School of Dentistry, Calif 92350, USA.
J Calif Dent Assoc. 2012 May;40(5):419-21.
This study investigated sterilization of used implant impression copings and healing abutments. Components were analyzed after contamination with Enterococcus foecalis, followed by multiple rounds of sterilization by both steam autoclave and Chemiclave protocols. The authors' results demonstrated that used components showed sterility equal to new components without any visible distortion. These data suggest that component resterilization and reuse may be justified or at least considered in clinical practice. Also, implications for cost savings in the placement of implants are advanced.
本研究调查了使用过的种植体印模帽和愈合基台的灭菌情况。在用粪肠球菌污染组件后进行分析,随后通过蒸汽高压灭菌器和化学灭菌程序进行多轮灭菌。作者的结果表明,使用过的组件显示出与新组件相同的无菌状态,且没有任何可见的变形。这些数据表明,组件的再次灭菌和再利用在临床实践中可能是合理的,或者至少值得考虑。此外,还提出了种植体植入成本节约方面的影响。