Suppr超能文献

一种新型无创设备用于检测 1 型糖尿病青少年夜间低血糖的原型研究——一项初步研究。

A prototype of a new noninvasive device to detect nocturnal hypoglycemia in adolescents with type 1 diabetes--a pilot study.

机构信息

Gili Medical Ltd., Migdal HaEmek, Israel.

出版信息

Diabetes Technol Ther. 2012 Aug;14(8):683-9. doi: 10.1089/dia.2012.0002. Epub 2012 Jun 12.

Abstract

BACKGROUND

Severe hypoglycemic events are a major consequence of tight diabetes control. Continuous glucose monitoring systems (CGMSs) were recently introduced in order to minimize the risk of hypoglycemia. However, the present CGMSs are invasive and costly and have been recently demonstrated to be intolerant for most children and adolescents. Hence there is a need for a simple, noninvasive, convenient, and inexpensive device to detect hypoglycemic events. The Gili Medical Hypoglycemia Non Invasive Monitoring System (GMHNIMS) (Gili Medical Ltd., Migdal HaEmek, Israel) has been currently developed for these purposes.

SUBJECTS AND METHODS

Ten patients 14-18 years old with type 1 diabetes for at least 1 year participated in a pilot study that was held at the Meyer Children's Hospital, Rambam Medical Center, Haifa, Israel. All patients were either treated by insulin pump or by multiple daily injections. The GMHNIMS was connected to the study subjects during three consecutive nights in an inpatient setting while they received their usual insulin regimen. The system is composed of four sensors (heart rate, perspiration, skin temperature, and tremor) that detect physiologic changes during hypoglycemia. In addition, each patient was connected to a real-time CGMS for 3 nights. When a hypoglycemic event was suspected clinically by the patient, a bedside capillary glucose was checked by a glucometer.

RESULTS

The system was found to be convenient without any disturbances to sleep quality. The sensitivity of the GMHNIMS for detection of true hypoglycemic events was 100% with specificity of 85.7%.

CONCLUSIONS

The new device showed high detection rates of nocturnal hypoglycemic events with an acceptable degree of false-positive readings. Being inexpensive and noninvasive, this device has the potential for routine use in insulin-treated patients.

摘要

背景

严重的低血糖事件是严格控制糖尿病的主要后果。为了最大限度地降低低血糖的风险,最近引入了连续血糖监测系统(CGMS)。然而,目前的 CGMS 具有侵入性和昂贵性,并且最近已经证明大多数儿童和青少年无法耐受。因此,需要一种简单、非侵入性、方便且廉价的设备来检测低血糖事件。Gili 医疗低血糖无创监测系统(GMHNIMS)(Gili Medical Ltd.,Migdal HaEmek,以色列)就是为此目的而开发的。

受试者和方法

10 名年龄在 14-18 岁、患有 1 型糖尿病至少 1 年的患者参加了在以色列海法拉宾医疗中心迈尔儿童医院进行的一项试点研究。所有患者均接受胰岛素泵或多次每日注射治疗。GMHNIMS 在住院期间连续三个晚上与研究对象连接,同时接受他们的常规胰岛素治疗方案。该系统由四个传感器(心率、出汗、皮肤温度和震颤)组成,可检测低血糖期间的生理变化。此外,每位患者还在三个晚上连接到实时 CGMS。当患者临床上怀疑发生低血糖事件时,使用血糖仪检查床边毛细血管血糖。

结果

该系统使用方便,不会对睡眠质量造成任何干扰。GMHNIMS 检测真正低血糖事件的灵敏度为 100%,特异性为 85.7%。

结论

该新设备对夜间低血糖事件的检测率较高,假阳性读数的程度可接受。这种设备价格低廉且非侵入性,有潜力在接受胰岛素治疗的患者中常规使用。