Division of Joint Replacement Surgery, Department of Orthopaedics and Traumatology, The University of Hong Kong, 216 New Clinical Building, Queen Mary Hospital, Pok Fu Lam, Hong Kong, SAR,
Knee Surg Sports Traumatol Arthrosc. 2013 Oct;21(10):2309-13. doi: 10.1007/s00167-012-2108-1. Epub 2012 Jun 26.
Femoral sizing in total knee replacement is important. Either undersizing or oversizing may result in deleterious effects to the clinical outcome after the surgery. There has been no study on the precision and accuracy of femoral sizing and the effect of measurement at different landmarks over the distal femur. This study assesses the intra-observer and inter-observer error of femoral sizing and identifies the effect of the placement site of the anterior referencing tool on femoral sizing.
Five investigators with different clinical experience measured the femoral size of 10 cadaveric specimens twice using three anterior referencing tool. The measurement of the femoral size was repeated at nine designated points on the anterior cortex of the cadaveric femora.
Excellent intraobserver and interobserver agreements were obtained using the three anterior referencing tools. When the size on which the majority agreed was regarded as the actual size of the specimen, measurement at the nine designated points on the anterior cortex showed a deviation from the actual size from 6.2 to 46.2 %. Placing the femoral sizer stylus at the middle and 2 cm above the proximal margin of the anterior femoral condyle yielded the highest precision and accuracy.
Regardless of the experience of the surgeons, measurement of the femoral size using the three anterior referencing tools is very accurate. Placing the stylus of the femoral sizer at the middle and 2 cm above the proximal margin of the anterior femoral condyle best reflects the actual size of the femur.
Experimental study.
全膝关节置换术中股骨的大小很重要。股骨尺寸过小或过大都可能对术后临床结果产生有害影响。目前还没有关于股骨尺寸测量的精度和准确性以及股骨远端不同标志点测量的研究。本研究评估了股骨尺寸测量的观察者内和观察者间误差,并确定了在前参照工具放置位置对股骨尺寸测量的影响。
5 名具有不同临床经验的研究人员使用 3 种前参照工具对 10 个尸体标本的股骨尺寸进行了两次测量。在尸体股骨前皮质的 9 个指定点重复测量股骨尺寸。
使用 3 种前参照工具获得了极好的观察者内和观察者间一致性。当以多数人同意的尺寸作为标本的实际尺寸时,在前皮质的 9 个指定点的测量值与实际尺寸的偏差为 6.2%至 46.2%。将股骨量规测针放置在前股骨髁近端边缘的中间和 2cm 处可获得最高的精度和准确性。
无论外科医生的经验如何,使用 3 种前参照工具测量股骨尺寸都非常准确。将股骨量规测针放置在前股骨髁近端边缘的中间和 2cm 处最能反映股骨的实际尺寸。