[An experimental study on rabbit liver by 3.0 Tesla magnetic resonance diffusion-weighted imaging].

作者信息

Dong Tian-Ming, Sun Xi-Jie, Xu Xian, Dai Jing, Zheng Wei, An Ning-Yu, Wang Xiao-Qing

机构信息

Department of Radiology, Chinese PLA General Hospital, Beijing 100853, China.

出版信息

Zhongguo Yi Xue Ke Xue Yuan Xue Bao. 2012 Feb;34(1):51-7.

Abstract

OBJECTIVE

To explore the methodology as well as the features, quantificational index, and reference standard of 3.0 Tesla magnetic resonance (MR) diffusion-weighted imaging (DWI) on the normal rabbit's liver.

METHODS

Twenty New Zealand white rabbits were enrolled and DWI was performed after anesthetics with multi-b values at 3.0 T MR scanner. Apparent diffusion coefficient (ADC) values as well as the difference between maximum and minimum ADC values, signal strength (SH), noise signal (SD), signal to noise ratio (SNR), and quality index (QI) were recorded and analyzed.

RESULTS

With b value increased, the ADC values decreased accordingly (P < 0.001). The difference between maximum and minimum ADC values with b = 1000 s/mm2 was the least (good stability), b = 600 s/mm2 was the second least, and b = 300 s/mm2 was greatest (bad stability). The SH decreased at the same time (P < 0.001), but the difference among DWI with b =600, 800, and 1000 s/mm2 was not statistically significant (P > 0.05). The SD decreased at the same time (P < 0.001), but the difference between DWI with b = 800 s/mm2 and b = 1000 s/mm2 was not statistically significant (P > 0.05). The SNR decreased at the same time (P < 0.001), but there were no significant differences between DWI with b = 600 s/mm2 and b = 800 s/mm2 or b = 600 s/mm2 and b = 1000 s/mm2 (P > 0.05). The SNR of DWI with b = 800 s/mm2 and b = 1000 s/mm2 was lower. The QI decreased at the same time (P < 0. 001) , but the difference between DWI with b = 800 s/mm2 and b = 1000 s/mm was not statistically significant (P > 0.05).

CONCLUSION

When 3.0 T MR DWI is applied for rabbit liver, it is better to use b = 600 s/mm2 for reducing scanning time and assuring better diffusion weights, quantity of images, and stability of ADC measurement.