An invasive but simple and accurate method for ascending aorta-femoral artery pulse wave velocity measurement.

作者信息

Kilic Harun, Yelgec Selcuk, Salih Orcan, Akdemir Ramazan, Karakurt Ozlem, Cagirci Goksel, Yeter Ekrem, Acikel Sadik, Dogan Mehmet, Arslantas Ugur, Baha Resat Mehmet

机构信息

Cardiology, Diskapi Yildirim Beyazit Education and Training Hospital, Ankara, Turkey.

出版信息

Blood Press. 2013 Feb;22(1):45-50. doi: 10.3109/08037051.2012.694255. Epub 2012 Jul 3.

Abstract

INTRODUCTION

Pulse wave velocity (PWV) is an emerging predictor in the assessment of cardiovascular risk in diseased and healthy populations. We suggest a novel method for the accurate measurement of PWV.

METHOD

PWV is calculated from pulse transit time using two separate pulse recordings over a known distance. 8F sheaths were placed in the right femoral arteries and routine coronary angiographies were performed with 5F diagnostic catheters. Ascending aorta pressures were measured with right diagnostic catheter tip in the ascending aorta and synchronous femoral artery pressures were measured with the sheath in the femoral artery. The distance between the two pressure sites was calculated as follows: total length of the right diagnostic catheter-length of the catheter outside the sheath-Sheath length.

RESULTS

We evaluated the PWV measured using the catheter method in 24 subjects. PWV correlated positively and independently with age (p = 0.004), coronary artery disease (p = 0.04), ascending aorta systolic pressure (p = 0.006), femoral artery systolic pressure (p = 0.008), ascending aorta pulse pressure (p = 0.003) and femoral artery pulse pressure (p = 0.04). In coronary artery disease patients, the mean PWV value was significantly higher than in patients with normal coronary arteries (12.61 ± 6.31 m/s vs 7.58 ± 2.26 m/s p = 0.04).

CONCLUSION

We describe a novel and accurate but invasive method for measurement of PWV. Our results may serve as a reference for non-invasive assessment of aorta-femoral artery PWV.

摘要

引言

脉搏波速度(PWV)是评估患病和健康人群心血管风险的一种新兴预测指标。我们提出了一种准确测量PWV的新方法。

方法

PWV通过在已知距离上使用两个单独的脉搏记录从脉搏传播时间计算得出。将8F鞘管置于右股动脉,并用5F诊断导管进行常规冠状动脉造影。使用位于升主动脉的右诊断导管尖端测量升主动脉压力,并用股动脉中的鞘管同步测量股动脉压力。两个压力位点之间的距离计算如下:右诊断导管的总长度-鞘管外导管的长度-鞘管长度。

结果

我们评估了24名受试者使用导管法测量的PWV。PWV与年龄呈正相关且独立相关(p = 0.004),与冠状动脉疾病(p = 0.04)、升主动脉收缩压(p = 0.006)、股动脉收缩压(p = 0.008)、升主动脉脉压(p = 0.003)和股动脉脉压(p = 0.04)相关。在冠状动脉疾病患者中,平均PWV值显著高于冠状动脉正常的患者(12.61±6.31米/秒对7.58±2.26米/秒,p = 0.04)。

结论

我们描述了一种新颖、准确但具有侵入性的PWV测量方法。我们的结果可为主动脉-股动脉PWV的无创评估提供参考。