Oak Farm House, Milverton, Skerries, Dublin, Ireland.
Med Phys. 2012 Jun;39(6):3019-25. doi: 10.1118/1.4711812.
UVR phototherapy cabins exist in a huge variety of configurations and geometries. An accurate dose model has been developed which can be used to examine the question of cabin design to investigate factors influencing dose on a patient.
This work extends the existing dose model to entire cabins and investigates how cabin and reflector geometry influence resultant dose.
The model predictions are in line with what is measured. It is found that the length and angle of the reflectors have a large influence on received dose.
The influence of cabin geometry is important in estimating patient dose, and the findings of this work are applicable to future cabin designs.
紫外线光疗舱有各种各样的配置和几何形状。我们开发了一种精确的剂量模型,可用于检查舱体设计问题,以研究影响患者剂量的因素。
这项工作将现有的剂量模型扩展到整个舱体,并研究舱体和反射器的几何形状如何影响最终剂量。
模型预测与实测结果相符。结果发现,反射器的长度和角度对接收剂量有很大影响。
在估算患者剂量时,舱体几何形状的影响很重要,这项工作的结果适用于未来的舱体设计。