Galgut P N, Waite I M
Periodontal Department, University College Hospital, London, UK.
Int Dent J. 1990 Dec;40(6):333-8.
This study compared clinical measurements using a manual periodontal probe and an electronic pressure probe using a 0.25 N force, and related them to measurement of the bone levels at the time of surgery. All measurements were taken from the cemento-enamel junction at specified sites. It was found that neither probing technique provided a means for accurately measuring bone level. In the presence of healed tissue, little difference was found between the two instruments; whereas the manual probe more closely approximated the actual level of bone in the presence of inflammation. It was therefore concluded that under the conditions of present study the manual probe depth measurements were the more applicable for routine clinical use.
本研究比较了使用手动牙周探针和施加0.25 N力的电子压力探针进行的临床测量,并将其与手术时的骨水平测量相关联。所有测量均在特定部位从牙骨质-釉质界处进行。结果发现,两种探测技术均未提供准确测量骨水平的方法。在愈合组织存在的情况下,两种仪器之间差异不大;而在存在炎症的情况下,手动探针更接近实际骨水平。因此得出结论,在本研究条件下,手动探针深度测量更适用于常规临床应用。