[用于哮喘和慢性阻塞性肺疾病吸入治疗装置选择的吸气流量测量]

[Measurement of inspiratory flow for the selection of the inhalation treatment device for asthma and COPD].

作者信息

Plavec Davor, Gluncić Tajana Jalusić, Gudelj Ivan, Mise Kornelija

机构信息

Djecja bolnica Srebrnjak, Zagreb.

出版信息

Lijec Vjesn. 2012 Mar-Apr;134(3-4):84-9.

Abstract

AIM

To assess the proportion of patients with asthma or chronic obstructive pulmonary disease (COPD) with significant bronchoobstruction who do not have inspiratory flows necessary for the adequate use of dry powder inhaler (DPI) devices Diskus and Turbuhaler.

PATIENTS AND METHODS

multicentre cross-sectional study that included 400 patients with asthma or COPD (with FEV1 < 60%) aged 5-91 years tested during exacerbation or usual control visit. All patients underwent lung function testing and measurement of inspiratory flows using In-check Dial.

RESULTS

Significantly lower proportion of patients didn't have inspiratory flows necessary for the regular adequate use of Diskus than for Turbuhaler (21.0% vs 87.0%, chi2 = 350.72, p < 0.0001). In patients tested during exacerbation these proportions were marginally greater. Regression analysis showed a weak association of inspiratory flows measured with different resistance settings with parameters of lung function (best association was seen between resistance setting for Diskus and PEF [%], r2 = 0.104).

CONCLUSIONS

Significant proportion of patients with asthma or COPD with significant bronchoobstruction do not exhibit satisfactory inspiratory flows for the use of dry powder inhaler (DPI) devices (Diskus < Turbuhaler). Spirometry is not to be used for selection of the device for drug application and In-check Dial should be used instead.

摘要

目的

评估患有哮喘或慢性阻塞性肺疾病(COPD)且存在严重支气管阻塞的患者中,没有足够吸气流量以充分使用干粉吸入器(DPI)装置(Diskus和都保)的患者比例。

患者与方法

多中心横断面研究,纳入400例年龄在5至91岁之间、患有哮喘或COPD(FEV1<60%)且在病情加重期或常规复诊时接受检查的患者。所有患者均接受肺功能测试,并使用In-check Dial测量吸气流量。

结果

与都保相比,没有足够吸气流量以常规充分使用Diskus的患者比例显著更低(21.0%对87.0%,χ2=350.72,p<0.0001)。在病情加重期接受检查的患者中,这些比例略高。回归分析显示,在不同阻力设置下测量的吸气流量与肺功能参数之间存在弱关联(在Diskus的阻力设置与呼气峰流速[%]之间观察到最佳关联,r2=0.104)。

结论

患有哮喘或COPD且存在严重支气管阻塞的患者中,有相当比例的患者没有表现出足以使用干粉吸入器(DPI)装置(Diskus<都保)的令人满意的吸气流量。肺活量测定不应被用于选择药物吸入装置,而应使用In-check Dial。