Kaulesar Sukul D M, Johannes E J, Marti R K, Klopper P J
Department of Traumatology, University Hospital Rotterdam Dijkzigt, Rotterdam, The Netherlands.
J Biomech. 1990;23(11):1115-21. doi: 10.1016/0021-9290(90)90004-m.
A number of screws commonly used for internal fixation in scaphoid bone fractures and nonunions are compared regarding biomechanical properties and clinical applicability. The experiments were carried out on models made of ash-wood, representing a reconstruction and fixation as is performed in a cortico-cancellous inlay bone graft for scaphoid non-union. For fixation use was made of 2.7 and 3.5 AO/ASIF cortical screws respectively, 4.0 AO/ASIF cancellous screws, Herbert screws, and a newly designed screw called the three components screw (D.K.S.). The models with implanted screws were tested for bending strength, tensile strength and torsion stability. No large differences between the various screws were found regarding the measured parameters, so that a small intra-osteal implant such as the Herbert screw and the D.K.S., which can be inserted easily and which gives a certain amount of interfragmentary compression, will be sufficient for osteosynthesis of the scaphoid bone. In case an intra-osteal implant is not available a single 3.5 AO/ASIF cortical screw, inserted following lag-screw principles, is recommended.
对一些常用于舟骨骨折和骨不连内固定的螺钉,在生物力学性能和临床适用性方面进行了比较。实验在由白蜡木制成的模型上进行,该模型代表了在皮质松质镶嵌骨移植治疗舟骨骨不连时所进行的重建和固定。固定分别使用了2.7和3.5的AO/ASIF皮质螺钉、4.0的AO/ASIF松质骨螺钉、Herbert螺钉以及一种新设计的名为三分体螺钉(D.K.S.)的螺钉。对植入螺钉的模型进行了抗弯强度、抗拉强度和扭转稳定性测试。在所测参数方面,各种螺钉之间未发现大的差异,因此,像Herbert螺钉和三分体螺钉这样的小型骨内植入物,因其易于插入并能提供一定量的骨折块间加压,足以用于舟骨的骨合成。如果没有骨内植入物,建议按照拉力螺钉原则插入一枚3.5的AO/ASIF皮质螺钉。