Department of Chemistry, Xixi Campus, Zhejiang University, Hangzhou 310028, China.
Department of Chemistry, Xixi Campus, Zhejiang University, Hangzhou 310028, China.
J Chromatogr A. 2012 Aug 17;1251:154-159. doi: 10.1016/j.chroma.2012.06.059. Epub 2012 Jun 25.
A simple ion chromatographic method for simultaneous detection of iodide and iodate in a single running was proposed, with columns packed with homemade functionalized polystyrene-divinylbenzene (PS-DVB) resins and column-switching technique. Homemade resins were functionalized with controllable amounts of quaternary ammonium groups. The low-capacity anion-exchange column and high-capacity anion-exchange column were prepared, due to the resins having different exchange capacities. With this method, iodide and iodate in povidone iodine solution were detected simultaneously in a short time with iodide being eluted off first. A series of standard solutions consisting of target anions of various concentrations from 0.01 mg/L to 100 mg/L were analyzed. Each anion exhibited satisfactory linearity, with correlation coefficient r ≥ 0.9990. The detection limits (LODs) for iodide and iodate obtained by injecting 100 μL of sample were 5.66 and 14.83 μg/L (S/N=3), respectively. A spiking study was performed with satisfactory recoveries between 101.2% and 100.6% for iodide and iodate.
提出了一种在单个运行中同时检测碘化物和碘酸盐的简单离子色谱法,使用填充有自制功能化聚苯乙烯-二乙烯基苯(PS-DVB)树脂的柱子和柱切换技术。自制树脂用可控数量的季铵基团功能化。由于树脂具有不同的交换能力,因此制备了低容量阴离子交换柱和高容量阴离子交换柱。使用该方法,可以在短时间内同时检测聚维酮碘溶液中的碘化物和碘酸盐,首先洗脱碘化物。分析了一系列浓度为 0.01 mg/L 至 100 mg/L 的目标阴离子的标准溶液。每个阴离子都表现出令人满意的线性,相关系数 r≥0.9990。通过注入 100 μL 样品,碘化物和碘酸盐的检测限(LOD)分别为 5.66 和 14.83 μg/L(S/N=3)。碘化物和碘酸盐的加标研究回收率在 101.2%至 100.6%之间,结果令人满意。