Smit Gerwin, Bongers Raoul M, Van der Sluis Corry K, Plettenburg Dick H
Delft Institute of Prosthetics and Orthotics, Department of Biomechanical Engineering, Delft University of Technology, the Netherlands.
J Rehabil Res Dev. 2012;49(4):523-34. doi: 10.1682/jrrd.2011.07.0125.
Quantitative data on the mechanical performance of upper-limb prostheses are very important in prostheses development and selection. The primary goal of this study was to objectively evaluate the mechanical performance of adult-size voluntary opening (VO) prosthetic terminal devices and select the best tested device. A second goal was to see whether VO devices have improved in the last two decades. Nine devices (four hooks and five hands) were quantitatively tested (Hosmer model 5XA hook, Hosmer Sierra 2 Load VO hook, RSL Steeper Carbon Gripper, Otto Bock model 10A60 hook, Becker Imperial hand, Hosmer Sierra VO hand, Hosmer Soft VO hand, RSL Steeper VO hand, Otto Bock VO hand). We measured the pinch forces, activation forces, cable displacements, mass, and opening span and calculated the work and hysteresis. We compared the results with data from 1987. Hooks required lower activation forces and delivered higher pinch forces than hands. The activation forces of several devices were very high. The pinch forces of all tested hands were too low. The Hosmer model 5XA hook with three bands was the best tested hook. The Hosmer Sierra VO hand was the best tested hand. We found no improvements in VO devices compared with the data from 1987.
上肢假肢机械性能的定量数据在假肢研发和选择中非常重要。本研究的主要目标是客观评估成人尺寸的主动开启(VO)假肢末端装置的机械性能,并选出测试效果最佳的装置。第二个目标是了解在过去二十年中VO装置是否有所改进。对九种装置(四个钩子和五只手)进行了定量测试(霍斯默5XA型钩子、霍斯默塞拉2型负载VO钩子、RSL Steeper碳夹、奥托博克10A60型钩子、贝克尔帝国之手、霍斯默塞拉VO手、霍斯默柔软VO手、RSL Steeper VO手、奥托博克VO手)。我们测量了捏力、启动力、缆绳位移、质量和开口跨度,并计算了功和滞后现象。我们将结果与1987年的数据进行了比较。与手相比,钩子所需的启动力更低,提供的捏力更高。几种装置的启动力非常高。所有测试手的捏力都过低。带有三条带的霍斯默5XA型钩子是测试效果最佳的钩子。霍斯默塞拉VO手是测试效果最佳的手。与1987年的数据相比,我们发现VO装置没有改进。