Li Yuanyuan, Henion Jack, Abbott Richard, Wang Phillip
Advion Bioanalytical Labs, Ithaca, NY, USA.
Bioanalysis. 2012 Jun;4(12):1445-56. doi: 10.4155/bio.12.111.
Direct analysis of dried blood spot (DBS) samples was investigated using a prototype semi-automated robotic device that allows the direct elution of sample spots from a DBS paper card to an online SPE cartridge. The eluted SPE samples were analyzed with high-performance LC-MS/MS.
A LLOQ of 0.01 ng/ml was achieved with a linear calibration range from 0.01 to 25 ng/ml. Optimal performance data were obtained from spotting the internal standard solution on the card before blood spotting. Internal standard addition from the system injector loop produced intra-assay inaccuracy of -9.0-7.3% and precision of 1.3-8.2%, and inter-assay inaccuracy of -3.5-3.9% and precision of 4.4-8.7%.
Results demonstrated the feasibility of a semi-automated online rapid direct elution method that avoids manual extraction for DBS sample analysis using the online DBS-SPE system coupled to LC-MS/MS.
使用一种原型半自动机器人设备对干血斑(DBS)样本进行直接分析,该设备可将DBS纸卡上的样本斑点直接洗脱至在线固相萃取(SPE)小柱。对洗脱后的SPE样本进行高效液相色谱-串联质谱(LC-MS/MS)分析。
实现了0.01 ng/ml的定量下限,线性校准范围为0.01至25 ng/ml。在血样点样前将内标溶液点在卡上可获得最佳性能数据。通过系统进样环添加内标,批内不准确度为-9.0-7.3%,精密度为1.3-8.2%;批间不准确度为-3.5-3.9%,精密度为4.4-8.7%。
结果证明了一种半自动在线快速直接洗脱方法的可行性,该方法避免了使用与LC-MS/MS联用的在线DBS-SPE系统对DBS样本分析进行手动提取。