Daller J R, Wong J, Brooks B D, McKee J S
Technology Resources, Baxter Healthcare Corporation, Round Lake, IL 60073, USA.
J Pharmacol Toxicol Methods. 2012 Nov-Dec;66(3):232-7. doi: 10.1016/j.vascn.2012.07.001. Epub 2012 Jul 13.
Commercial whole-body plethysmography systems used to evaluate the anti-tussive potential of drugs employ sophisticated technology, but these systems may be cost prohibitive for some laboratories. The present study describes an alternative, inexpensive system for evaluating the tussive and anti-tussive potential of drugs in conscious, unrestrained guinea pigs.
The system is composed of a transparent small animal anesthesia induction box fitted with a microphone, a camera and a pneumotachometer to simultaneously capture audio, video, air flow and air pressure in real time. Data acquisition and analysis was performed using free software for audio and video, and a research pneumotach system for flow and pressure. System suitability testing was performed by exposing conscious, unrestrained guinea pigs to nebulized aqueous solutions of a selective agonist for TRPV1 (citric acid) or a selective agonist for TRPA1 (AITC), with or without pre-treatment with a selective antagonist for TRPV1 (BCTC) or a selective antagonist for TRPA1 (HC-030031).
The system easily discerned coughs from other respiratory events like sneezes. System suitability test results are as follows: AITC caused 10.7 (SEM=1.4592) coughs vs. 5.8 (SEM=1.6553) when pre-treated with HC-030031 (P<0.05). Citric acid caused 12.4 (SEM=1.4697) coughs vs. 3.2 (SEM=1.3928) when pre-treated with BCTC (P<0.002).
We have described in detail an inexpensive system for evaluating the tussive and anti-tussive potential of chemicals in conscious, unrestrained guinea pigs. Suitability testing indicates that the system is comparable to a commercial whole-body plethysmography system for detecting and differentiating between coughs and sneezes. This system may provide some investigators a cost-conscious alternative to more expensive commercial whole-body plethysmography systems.
用于评估药物镇咳潜力的商用全身体积描记系统采用了复杂的技术,但这些系统对一些实验室来说成本可能过高。本研究描述了一种用于评估清醒、不受束缚的豚鼠药物咳嗽和镇咳潜力的低成本替代系统。
该系统由一个透明的小动物麻醉诱导箱组成,箱内装有麦克风、摄像头和呼吸流速计,可实时同时捕捉音频、视频、气流和气压。使用免费的音频和视频软件以及用于流量和压力的研究型呼吸流速系统进行数据采集和分析。通过将清醒、不受束缚的豚鼠暴露于TRPV1的选择性激动剂(柠檬酸)或TRPA1的选择性激动剂(AITC)的雾化水溶液中,无论是否预先用TRPV1的选择性拮抗剂(BCTC)或TRPA1的选择性拮抗剂(HC - 030031)进行预处理,来进行系统适用性测试。
该系统能够轻松地将咳嗽与打喷嚏等其他呼吸事件区分开来。系统适用性测试结果如下:AITC引起10.7次(标准误 = 1.4592)咳嗽,而预先用HC - 030031处理后为5.8次(标准误 = 1.6553)(P < 0.05)。柠檬酸引起12.4次(标准误 = 1.4697)咳嗽,而预先用BCTC处理后为3.2次(标准误 = 1.3928)(P < 0.002)。
我们详细描述了一种用于评估清醒、不受束缚的豚鼠中化学物质咳嗽和镇咳潜力的低成本系统。适用性测试表明,该系统在检测和区分咳嗽与打喷嚏方面与商用全身体积描记系统相当。该系统可能为一些研究人员提供一种比更昂贵的商用全身体积描记系统更具成本意识的替代方案。