Lima Pedro O P, Oliveira Rodrigo R, Moura Filho Alberto G, Raposo Maria C F, Costa Leonardo O P, Laurentino Glória E C
Physical Therapy Department, Universidade Federal do Ceará, Fortaleza, CE, Brazil.
Rev Bras Fisioter. 2012 Sep-Oct;16(5):389-95. doi: 10.1590/s1413-35552012005000038. Epub 2012 Jul 24.
The Pressure biofeedback unit (PBU) is an assessment tool used in clinical practice and research aimed to indirectly analyze the transversus abdominis (TrA) muscle activity. The concurrent validity of the PBU in a clinically relevant sample is still unclear.
The purpose of this study was to evaluate the concurrent validity and diagnostic accuracy of the PBU in measuring TrA muscle activity in patients with chronic nonspecific low back pain.
This study was performed using a validation, cross-sectional design. Fifty patients with chronic nonspecific low back pain were recruited for this study. To test the concurrent validity both PBU measures (index test) and superficial electromyographic measures (reference-standard test) were compared and collected by a physical therapist in a series of voluntary contraction maneuvers of TrA muscle.
Participants were on average 22 years old, weighed 63.7 kilos, 1.70 meters height and mean low back pain duration was 1.9 years. It was observed a weak and non-significant Phi coefficient (r=0.2, p<0.20). With regards to diagnostic accuracy tests, our results suggest a low sensitivity (60%) and specificity (60%) of the PBU. The positive predictive value was high (0.8) and negative predictive value was low (0.2).
Concurrent validity of the PBU in measuring TrA muscle activity in patients with chronic nonspecific low back pain is poor given the low correlation and diagnostic accuracy with superficial EMG.
压力生物反馈仪(PBU)是一种用于临床实践和研究的评估工具,旨在间接分析腹横肌(TrA)的肌肉活动。PBU在临床相关样本中的同时效度仍不明确。
本研究的目的是评估PBU在测量慢性非特异性下腰痛患者TrA肌肉活动时的同时效度和诊断准确性。
本研究采用验证性横断面设计。招募了50名慢性非特异性下腰痛患者参与本研究。为了测试同时效度,在一系列TrA肌肉的自主收缩动作中,由一名物理治疗师比较并收集PBU测量值(指标测试)和表面肌电图测量值(参考标准测试)。
参与者平均年龄22岁,体重63.7公斤,身高1.70米,下腰痛平均持续时间为1.9年。观察到Phi系数较弱且无统计学意义(r = 0.2,p < 0.20)。关于诊断准确性测试,我们的结果表明PBU的敏感性较低(60%),特异性较低(60%)。阳性预测值较高(0.8),阴性预测值较低(0.2)。
鉴于与表面肌电图的低相关性和诊断准确性,PBU在测量慢性非特异性下腰痛患者TrA肌肉活动时的同时效度较差。