Validation of a novel spinal posture monitor: comparison with digital videofluoroscopy.

机构信息

Physiotherapy Department, University of Limerick, Limerick, Ireland.

出版信息

Eur Spine J. 2012 Dec;21(12):2633-9. doi: 10.1007/s00586-012-2440-7. Epub 2012 Jul 27.

Abstract

PURPOSE

A novel, minimally invasive posture monitor which can monitor lumbar postures outside the laboratory has demonstrated excellent reliability, as well as concurrent validity compared to a surface marker-based motion analysis system. However, it is unclear if this device reflects underlying vertebral motion.

METHODS

Twelve participants performed full range sagittal plane lumbo-pelvic movements during sitting and standing. Their posture was measured simultaneously using both this device (BodyGuard™) and digital videofluoroscopy.

RESULTS

Strong correlations were observed between the two methods (all r (s) > 0.88). Similarly, the coefficients of determination were high (all r (2) > 0.78). The maximum mean difference between the measures was located in the mid-range of motion and was approximately 3.4° in sitting and 3.9° in standing.

CONCLUSION

The BodyGuard™ appears to be a valid method for analysing vertebral motion in the sagittal plane and is a promising tool for long-term monitoring of spinal postures in laboratory and clinical settings in people with low back pain.

摘要

目的

一种新型的微创姿势监测器,可在实验室外监测腰椎姿势,与基于表面标记的运动分析系统相比,具有出色的可靠性和同时效度。然而,目前尚不清楚该设备是否反映了潜在的椎体运动。

方法

12 名参与者在坐姿和站立位时进行了全范围矢状面腰骨盆运动。使用该设备(BodyGuard™)和数字透视荧光检查同时测量他们的姿势。

结果

两种方法之间观察到很强的相关性(所有 r(s)>0.88)。同样,确定系数也很高(所有 r(2)>0.78)。在运动的中间范围内,测量值之间的最大平均差异约为 3.4°,在坐姿中为 3.9°,在站立位中为 3.9°。

结论

BodyGuard™ 似乎是一种分析矢状面椎体运动的有效方法,是一种有前途的工具,可在实验室和临床环境中对腰痛患者的脊柱姿势进行长期监测。