This study aimed to identify the optimal concentration, injection rate, and total volume of contrast medium (CM) for evaluating the hemodynamics of a hepatocellular carcinoma (HCC) structure model of diameter 35 mm, using multi-level dynamic computed tomography (M-LDCT) with 64 detector rows. A tube was inserted in the model as a simulated vessel. Five CM concentrations were used: non-diluted, 2-, 3-, 6-, and 9-fold diluted. Five regions of interest were placed within the HCC structure model. Time-density curves were created for CM injection rates of 1, 2, and 3 ml/s for 10 s, and for a total volume of 10 ml, followed by saline injection at 1 ml/s. M-LDCT maximum intensity projection images were evaluated by four appraisers using a three-point scale (excellent, 2; good, 1; poor, 0). There was no significant difference between maximum CT values at 2 ml/s for 10 s and those at 3 ml/s; these values were both greater than those at 1 ml/s. The duration of the peak was maintained for longer at 3 ml/s for 10 s (5.2 ± 2.3 s) than at 2 ml/s (3.6 ± 0.9 s). Maximum CT values at 2 ml/s of a total volume of 10 ml were greater than those at 3 ml/s. The highest scores of 7 and 8 were found at 2 and 3 ml/s for 10 s, using 2-, 3-, or 6-fold diluted CM. The most appropriate CM rate for evaluating hemodynamics was 2 ml/s for 10 s, using 2-, 3-, or 6-fold diluted CM.
本研究旨在通过使用64排探测器的多层动态计算机断层扫描(M-LDCT),确定用于评估直径35毫米的肝细胞癌(HCC)结构模型血流动力学的造影剂(CM)的最佳浓度、注射速率和总体积。在模型中插入一根管子作为模拟血管。使用了五种CM浓度:未稀释、2倍、3倍稀释、6倍稀释和9倍稀释。在HCC结构模型内放置了五个感兴趣区域。针对1、2和3毫升/秒的CM注射速率持续10秒、总体积为10毫升的情况创建时间-密度曲线,随后以1毫升/秒的速率注射生理盐水。四位评估者使用三点量表(优秀,2分;良好,1分;差,0分)对M-LDCT最大强度投影图像进行评估。10秒内2毫升/秒时的最大CT值与3毫升/秒时的最大CT值之间无显著差异;这两个值均大于1毫升/秒时的值。10秒内3毫升/秒时峰值持续时间(持续5.2±2.3秒)比2毫升/秒时(持续3.6±0.9秒)更长。总体积为10毫升时,2毫升/秒时的最大CT值大于3毫升/秒时的值。使用2倍、3倍或6倍稀释的CM时,10秒内2毫升/秒和3毫升/秒时的评分最高,分别为7分和8分。评估血流动力学最合适的CM注射速率是10秒内2毫升/秒,使用2倍、3倍或6倍稀释的CM。