PURPOSE

Tibial nail interlocking screw failure often occurs during delayed fracture consolidation or at early weight bearing of nailed unstable fractures, in general when high implant stress could not be reduced by other means. Is there a biomechanical improvement in long-term performance of angle stable locking screws compared to conventional locking screws for distal locking of intramedullary tibial nails?

METHODS

Surrogate bones of human tibiae were cut in the distal third and distal locking of the 10 mm intramedullary tibial nail was performed with either two angle stable locking screws or two conventional locking screws in the mediolateral plane. Six specimens per group were mechanically tested under quasi-static and cyclic axial loading with constantly increasing force.

RESULTS

Angle stable locking screw constructs exhibited significantly higher stiffness values (7,809 N/mm ± 647, mean ± SD) than conventional locking screw constructs (6,614 N/mm ± 859, p = 0.025). Angle stable locking screw constructs provided a longer fatigue life, expressed in a significantly higher number of cycles to failure (187,200 ± 18,100) compared to conventional locking screw constructs (128,700 ± 7,000, p = 0.004).

CONCLUSION

Fatigue performance of locking screws can be ameliorated by the use of angle stable locking screws, being especially important if the nail acts as load carrier and an improved stability during fracture healing is needed.

摘要

目的

在延迟骨折愈合或钉不稳定骨折早期负重时,胫骨钉锁钉常常发生失效,通常是在高植入物应力无法通过其他方式降低时。与传统锁定螺钉相比,角稳定锁定螺钉在髓内胫骨钉远端锁定中的长期性能是否具有生物力学改善?

方法

用人胫骨的替代骨在远端三分之一处切割,并在横平面中用两根角稳定锁定螺钉或两根传统锁定螺钉进行 10mm 髓内胫骨钉的远端锁定。每组 6 个标本在准静态和循环轴向加载下进行机械测试,施加的力逐渐增加。

结果

角稳定锁定螺钉结构的刚度值(7809N/mm±647,平均值±标准差)明显高于传统锁定螺钉结构(6614N/mm±859,p=0.025)。角稳定锁定螺钉结构的疲劳寿命更长,失效循环数(1872000±18100)明显高于传统锁定螺钉结构(1287000±7000,p=0.004)。

结论

使用角稳定锁定螺钉可以改善锁定螺钉的疲劳性能,如果钉作为载荷载体且需要改善骨折愈合期间的稳定性,则尤其重要。