I.N.F.N. Section of Milan, Via Celoria 16, 20133 Milano, Italy.
Phys Med. 2013 Nov;29(6):656-61. doi: 10.1016/j.ejmp.2012.07.005. Epub 2012 Aug 14.
The purpose of this work is to investigate the capability of the FLUKA Monte Carlo (MC) code to simulate the Elekta Leksell Gamma Knife Perfexion (LGK-PFX) and reproduce the Treatment Planning System (TPS) Leksell GammaPlan version 8.2 (LGP) dose calculations for the case of a water equivalent phantom target. Thanks to the collaboration with Elekta Instruments AB, the collimation system geometry, the source positions and all the involved material have been simulated in detail. The relative linear dose distribution along the three coordinate axes, for each collimator size, and the Relative Output Factors (ROF) have been investigated. The simulation has been validated comparing simulated linear dose profiles with measurements performed with EBT radiochromic films. The acceptance criterion between experimental data and FLUKA results is based on the gamma index (GI) method. The FLUKA MC calculation for the ROF provided the values of 0.920 for the 8 mm collimators and 0.800 for the 4 mm collimators. These values are in good agreement with the Elekta reference data of 0.924 and 0.805 respectively. The percentage difference between calculated and reference values for the ROF is under 1% and within the FLUKA uncertainty. Also the simulated relative dose profiles show a good agreement with the LGP calculation expressed by means of the gamma index method. This established accuracy proves that FLUKA is a suitable and powerful tool in order to reproduce successfully the LGP calculations for the homogeneous media.
这项工作的目的是研究 FLUKA 蒙特卡罗(MC)代码模拟 Elekta Leksell Gamma Knife Perfexion(LGK-PFX)并再现治疗计划系统(TPS)Leksell GammaPlan 版本 8.2(LGP)剂量计算的能力,针对水等效 phantom 靶的情况。由于与 Elekta Instruments AB 的合作,准直系统几何形状、源位置和所有涉及的材料都得到了详细模拟。研究了每个准直器尺寸的三个坐标轴向的相对线性剂量分布以及相对输出因子(ROF)。通过将模拟线性剂量分布与使用 EBT 放射自显影胶片进行的测量进行比较,对模拟进行了验证。实验数据和 FLUKA 结果之间的接受标准基于伽马指数(GI)方法。FLUKA MC 计算的 ROF 值为 8mm 准直器的 0.920 和 4mm 准直器的 0.800。这些值与 Elekta 的参考数据分别为 0.924 和 0.805 非常吻合。ROF 的计算值与参考值之间的百分比差异小于 1%,且在 FLUKA 不确定性范围内。模拟的相对剂量分布也与通过伽马指数方法表示的 LGP 计算具有良好的一致性。这种已确立的准确性证明了 FLUKA 是一种合适且强大的工具,可成功再现均质介质的 LGP 计算。