Department of Internal Medicine Clinical Laboratory, School of Dentistry, Health Sciences University of Hokkaido, Ishikari-Tobetsu, Hokkaido, Japan.
J Thromb Haemost. 2012 Nov;10(11):2338-43. doi: 10.1111/j.1538-7836.2012.04906.x.
Lupus anticoagulant (LA) is an antibody that interferes with phospholipid-dependent coagulation reactions. Activated partial thromboplastin time (APTT) is widely used as a test for LA screening. APTT reagents are composed of activators, such as silica or ellagic acid, and phospholipids, and APTT reagents with silica are recommended for LA screening because of greater sensitivity. However, the effects of activators on LA activity have not been adequately investigated.
In this study, we examined whether an ellagic acid-based reagent was highly sensitive to LA in a low phospholipid condition and useful for LA screening.
Silica-based (SL) and ellagic acid-based (EA) reagents were prepared in-house with the same composition and concentration of phospholipids, while the commercial APTT reagents APTT-SLA (SLA), Actin FSL (FSL), APTT-SP (SP) and PTT-LA (PTT) were also included in the study.
The normal reference ranges for SL and EA were 30.1-47.0 and 28.0-40.2 s, respectively, while the cut-off index values for circulating anticoagulant activity (ICA) calculated from the results obtained with SL, EA, SLA, FSL, SP and PTT were 12.9, 11.5, 13.2, 15.6, 14.3 and 14.0, respectively. The sensitivity of those reagents based on those cut-off values was 91%, 96%, 68%, 46%, 91% and 86%, respectively.
Our results showed that the ellagic acid-based reagent was more sensitive to LA than silica-based reagents in a low phospholipid condition and had adequate sensitivity to detect LA. We concluded that the sensitivity of APTT reagents for LA is dependent on phospholipid concentration and not the activator.
狼疮抗凝物(LA)是一种干扰磷脂依赖性凝血反应的抗体。活化部分凝血活酶时间(APTT)被广泛用作 LA 筛选的检测方法。APTT 试剂由激活剂(如二氧化硅或鞣花酸)和磷脂组成,由于具有更高的敏感性,推荐使用含有二氧化硅的 APTT 试剂进行 LA 筛选。然而,激活剂对 LA 活性的影响尚未得到充分研究。
本研究旨在探讨鞣花酸基试剂在低磷脂条件下对 LA 是否具有高度敏感性,以及是否可用于 LA 筛选。
采用相同组成和磷脂浓度自制了基于二氧化硅(SL)和鞣花酸(EA)的试剂,并纳入了商业 APTT 试剂 APTT-SLA(SLA)、Actin FSL(FSL)、APTT-SP(SP)和 PTT-LA(PTT)进行研究。
SL 和 EA 的正常参考范围分别为 30.1-47.0 和 28.0-40.2 s,而根据 SL、EA、SLA、FSL、SP 和 PTT 检测结果计算出的循环抗凝活性(ICA)的截断指数值分别为 12.9、11.5、13.2、15.6、14.3 和 14.0。基于这些截断值,这些试剂的敏感性分别为 91%、96%、68%、46%、91%和 86%。
我们的结果表明,在低磷脂条件下,鞣花酸基试剂比二氧化硅基试剂对 LA 更敏感,并且具有足够的敏感性来检测 LA。我们得出结论,APTT 试剂对 LA 的敏感性取决于磷脂浓度,而不是激活剂。