Compare the accuracy and reliability of fetal heart rate identification from maternal abdominal fetal electrocardiogram signals (ECG) and Doppler ultrasound with a fetal scalp electrode.
Prospective open method equivalence study.
Three urban teaching hospitals in the Northeast United States.
75 women with normal pregnancies in labor at >37 weeks of gestation.
Three fetal heart rate detection methods were used simultaneously in 75 parturients. The fetal scalp electrode was the standard against which abdominal fetal ECG and ultrasound were judged.
The positive percent agreement with the fetal scalp electrode indicated reliability. Bland-Altman analysis determined accuracy. The confusion rate indicated how frequently the devices tracked the maternal heart rate.
Positive percent agreement was 81.7 and 73% for the abdominal fetal ECG and ultrasound, respectively (p = 0.002). The abdominal fetal ECG had a lower root mean square error than ultrasound (5.2 vs. 10.6 bpm, p < 0.001). The confusion rate for ultrasound was 20-fold higher than for abdominal ECG (8.9 vs. 0.4%, respectively, p < 0.001).
Compared with the fetal scalp electrode, fetal heart rate detection using abdominal fetal ECG was more reliable and accurate than ultrasound, and abdominal fetal ECG was less likely than ultrasound to display the maternal heart rate in place of the fetal heart rate.
比较从母体腹部胎儿心电图(ECG)和多普勒超声信号与胎儿头皮电极获得的胎儿心率识别的准确性和可靠性。
前瞻性开放性方法等效性研究。
美国东北部的三家城市教学医院。
75 名正常妊娠分娩的孕妇,妊娠时间超过 37 周。
75 名产妇同时使用三种胎儿心率检测方法。胎儿头皮电极是判断腹部胎儿 ECG 和超声的标准。
与胎儿头皮电极的阳性百分率表示可靠性。Bland-Altman 分析确定准确性。混淆率表明设备跟踪母体心率的频率。
腹部胎儿 ECG 和超声的阳性百分率分别为 81.7%和 73%(p=0.002)。腹部胎儿 ECG 的均方根误差低于超声(5.2 与 10.6 bpm,p<0.001)。超声的混淆率比腹部 ECG 高 20 倍(分别为 8.9%和 0.4%,p<0.001)。
与胎儿头皮电极相比,使用腹部胎儿 ECG 检测胎儿心率的准确性和可靠性更高,并且腹部胎儿 ECG 比超声更不可能显示母体心率代替胎儿心率。