分娩期间胎儿心率监测:三种方法的比较

Monitoring Fetal Heart Rate during Labor: A Comparison of Three Methods.

作者信息

Euliano Tammy Y, Darmanjian Shalom, Nguyen Minh Tam, Busowski John D, Euliano Neil, Gregg Anthony R

机构信息

Department of Anesthesiology, University of Florida College of Medicine, Gainesville, FL, USA; Department of Obstetrics and Gynecology, University of Florida College of Medicine, Gainesville, FL, USA.

OBMedical, Jonesville, FL, USA.

出版信息

J Pregnancy. 2017;2017:8529816. doi: 10.1155/2017/8529816. Epub 2017 Mar 14.

Abstract

The purpose of the study was to compare the accuracy of a noninvasive fetal heart rate monitor with that of ultrasound, using a fetal scalp electrode as the gold standard, in laboring women of varying body habitus, throughout labor and delivery. Laboring women requiring fetal scalp electrode were monitored simultaneously with the investigational device (noninvasive fetal ECG), ultrasound, and fetal scalp electrode. An algorithm extracted the fetal heart rate from the noninvasive fetal ECG signal. Each noninvasive device recording was compared with fetal scalp electrode with regard to reliability by positive percent agreement and accuracy by root mean squared error. Seventy-one women were included in this analysis. Positive percent agreement was 83.4 ± 15.4% for noninvasive fetal ECG and 62.4 ± 26.7% for ultrasound. The root mean squared error compared with fetal scalp electrode-derived fetal heart rate was 4.8 ± 2.0 bpm for noninvasive fetal ECG and 14.3 ± 8.2 bpm for ultrasound. The superiority of noninvasive fetal ECG was maintained for stages 1 and 2 of labor and increases in body mass index. Compared with fetal scalp electrode-derived fetal heart rate, noninvasive fetal ECG is more accurate and reliable than ultrasound for intrapartum monitoring for stages 1 and 2 of labor and is less affected by increasing maternal body mass index. This confirms the results of other workers in this field.

摘要

本研究的目的是,在不同体型的分娩妇女整个分娩过程中,以胎儿头皮电极作为金标准,比较一种无创胎儿心率监测仪与超声的准确性。对需要胎儿头皮电极的分娩妇女,同时使用研究设备(无创胎儿心电图)、超声和胎儿头皮电极进行监测。一种算法从无创胎儿心电图信号中提取胎儿心率。通过阳性百分比一致性比较每种无创设备记录与胎儿头皮电极的可靠性,并通过均方根误差比较准确性。本分析纳入了71名妇女。无创胎儿心电图的阳性百分比一致性为83.4±15.4%,超声为62.4±26.7%。与胎儿头皮电极得出的胎儿心率相比,无创胎儿心电图的均方根误差为4.8±2.0次/分钟,超声为14.3±8.2次/分钟。在分娩第1和第2阶段以及体重指数增加时,无创胎儿心电图的优势依然存在。与胎儿头皮电极得出的胎儿心率相比,无创胎儿心电图在分娩第1和第2阶段的产时监测中比超声更准确、更可靠,且受孕妇体重指数增加的影响较小。这证实了该领域其他研究人员的结果。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/16cd/5368359/ae3db4078384/JP2017-8529816.001.jpg