Device life of the Provox Vega Indwelling voice prosthesis is as yet untested outside Europe. The current study examined device life and reasons for replacement within an Australian clinical setting. Twenty-three participants were monitored for device life and reasons for replacement. Main outcome measure was days to failure of initial device. Average device life and reasons for replacement were secondary measures. Initial device life data revealed 67 % had functioning devices at 3 months, 52 % at 6 months and 29 % at 12 months. Average device life was 207 days (median of 222). The majority of devices (97 %) failed due to leakage through the prosthesis. The Provox Vega Indwelling voice prosthesis had favourable device life in this cohort of patients and in comparison to European data. Reasons for replacement were consistent with international literature.

摘要

该植入式 Provox Vega 声门上型言语假体在欧洲以外地区的使用寿命尚未经过检验。本研究在澳大利亚临床环境中对设备的使用寿命和更换原因进行了检查。23 名参与者被监测设备的使用寿命和更换原因。主要观察指标是初始设备失效的天数。平均设备寿命和更换原因是次要观察指标。初始设备寿命数据显示,67%的患者在 3 个月时有功能正常的设备,52%的患者在 6 个月时有功能正常的设备,29%的患者在 12 个月时有功能正常的设备。平均设备寿命为 207 天(中位数为 222 天)。大多数设备(97%)因假体渗漏而失效。与欧洲数据相比,该植入式 Provox Vega 声门上型言语假体在这组患者中有良好的设备使用寿命。更换的原因与国际文献一致。