Suppr超能文献

临床质谱仪外周采样位点的频率响应

Frequency response of the peripheral sampling sites of a clinical mass spectrometer.

作者信息

Carlon G C, Kopec I C, Miodownik S, Ray C

机构信息

Department of Anesthesiology and Critical Care Medicine, Memorial Sloan-Kettering Cancer Center, New York, New York 10021.

出版信息

Anesthesiology. 1990 Jan;72(1):187-90. doi: 10.1097/00000542-199001000-00027.

Abstract

Mass spectrometers are used in ICUs and ORs to measure the concentration of medical and anesthetic gases gathered from multiple sites. This investigation was designed to determine the accuracy of a clinical system, which included 12 ICU bedside stations monitored by a medical mass spectrometer (Perkin-Elmer RMS III, Pomona, CA). Each site station was connected to the analyzing unit via two Teflon tubes, one permanently installed, 30-m long, and the second disposable, 2.4-m long. A gas mixture containing 95% O2 and 5% CO2, alternating with room air, was delivered to a solenoid valve and from there to the connecting tubes. Gas flow-rate, delay time, rise time, and peak and trough concentrations were determined for each gas at solenoid cycling frequencies of 25, 50, and 100/min. After the first set of measurements, the 30-m tubes were thoroughly cleaned and all measurements repeated. In addition, the authors also measured CO2 delay and rise times when the gas was delivered to the mass spectrometer through an unused 30-m tube or a new 2.4-m tube. Gas flow-rate increased from 143 +/- 12 ml/min (mean +/- SD) to 238 +/- 9 ml/min after the tubes were cleaned. Delay time was identical for all gases at all solenoid cycling rates but decreased significantly (P less than 0.05), from 11.5 +/- 0.3 to 4.8 +/- 0.7 s after the ceiling tubes were cleaned. As solenoid valve rate increased, the difference between measured and actual gas concentration increased. The lowest accuracy was 63.6 +/- 2.1%, for CO2 at 100 cycles/min.(ABSTRACT TRUNCATED AT 250 WORDS)

摘要

质谱仪用于重症监护病房(ICU)和手术室,以测量从多个部位收集的医用气体和麻醉气体的浓度。本研究旨在确定一种临床系统的准确性,该系统包括12个由医用质谱仪(Perkin-Elmer RMS III,加利福尼亚州波莫纳)监测的ICU床边工作站。每个站点通过两根聚四氟乙烯管连接到分析单元,一根是永久安装的30米长的管子,另一根是一次性的2.4米长的管子。一种含有95%氧气和5%二氧化碳的气体混合物与室内空气交替输送到电磁阀,然后从那里输送到连接管。在电磁阀循环频率为25、50和100次/分钟时,测定每种气体的气体流速、延迟时间、上升时间以及峰值和谷值浓度。在进行第一组测量后,对30米长的管子进行彻底清洁并重复所有测量。此外,作者还测量了气体通过未使用的30米长管子或新的2.4米长管子输送到质谱仪时的二氧化碳延迟时间和上升时间。管子清洁后,气体流速从143±12毫升/分钟(平均值±标准差)增加到238±9毫升/分钟。在所有电磁阀循环速率下,所有气体的延迟时间相同,但在天花板管子清洁后,延迟时间显著缩短(P<0.05),从11.5±0.3秒降至4.8±0.7秒。随着电磁阀速率的增加,测量的气体浓度与实际气体浓度之间的差异增大。在100次/分钟时,二氧化碳的最低准确度为63.6±2.1%。(摘要截短于250字)

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验