Suppr超能文献

下颌前徙术中双侧矢状劈开下颌支截骨改良接骨板技术的体外生物力学评估

In vitro biomechanical evaluation of modified plating techniques for bilateral sagittal split ramus osteotomy in mandibular advancement.

作者信息

Sener Ismail, Arıcı Selim, Bereket Cihan, Tek Mustafa

机构信息

Department of Oral and Maxillofacial Surgery, Faculty of Dentistry, Ondokuz Mayis University, Samsun, Turkey.

出版信息

J Craniofac Surg. 2012 Sep;23(5):1573-6. doi: 10.1097/SCS.0b013e31826100ec.

Abstract

The aim of this study was to compare the biomechanical properties of modified plate techniques using a newly designed three-dimensional test model after sagittal split ramus osteotomy. Fifteen synthetic polyurethane mandibular replicas were used and divided into 3 groups. Self-tapping 6-mm titanium screws with an outer diameter of 2.0 mm and private designed plates produced by 316-L stainless steel were used in the study. After the osteotomy, the distal part was advanced by 5 mm. The jaw models were placed to the three-dimensional test model designed by the authors, and the Instron Lloyd LRX device was used to apply a double-sided tensile force from the mandible angulus region of each group. Resistance forces that caused a displacement of 1.5 and 3 mm were recorded with the Instron program. The mean (SD) of the 3 groups were calculated using analysis of variance and the Tukey test. The results were compared statistically, with values of P < 0.05 determined as statistically significant. On the basis of the results of the Tukey comparison within the groups, there was a statistically significant difference between groups 1 and 2 and between groups 1 and 3 at both 1.5- and 3-mm displacement (P < 0.05). No significant difference was observed between groups 2 and 3 (P > 0.05). This test model was able to identify the most appropriate plate type of 3 different modified plate techniques after sagittal split ramus osteotomy.

摘要

本研究的目的是在矢状劈开下颌支截骨术后,使用新设计的三维测试模型比较改良接骨板技术的生物力学特性。使用了15个合成聚氨酯下颌骨复制模型,并将其分为3组。本研究使用了外径为2.0 mm的自攻型6 mm钛螺钉以及由316-L不锈钢制作的定制接骨板。截骨术后,将远心端向前推进5 mm。将下颌骨模型放置于作者设计的三维测试模型中,并使用英斯特朗Lloyd LRX设备从每组的下颌角区域施加双侧拉力。使用英斯特朗程序记录导致1.5 mm和3 mm位移的阻力。使用方差分析和Tukey检验计算3组的均值(标准差)。对结果进行统计学比较,P < 0.05的值被确定为具有统计学显著性。根据组内Tukey比较的结果,在1.5 mm和3 mm位移时,第1组和第2组之间以及第1组和第3组之间存在统计学显著差异(P < 0.05)。第2组和第3组之间未观察到显著差异(P > 0.05)。该测试模型能够识别矢状劈开下颌支截骨术后3种不同改良接骨板技术中最合适的接骨板类型。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验