Suppr超能文献

不同测量技术测量角膜散光的可比性和可重复性。

Comparability and repeatability of corneal astigmatism measurements using different measurement technologies.

机构信息

University Eye Clinic Maastricht, Maastricht University Medical Centre, The Netherlands.

出版信息

J Cataract Refract Surg. 2012 Oct;38(10):1764-70. doi: 10.1016/j.jcrs.2012.05.036.

Abstract

PURPOSE

To determine the comparability and repeatability of corneal astigmatism measurements obtained with different devices and determine the interobserver variability of a new automated keratometer.

SETTING

University Eye Clinic Maastricht, the Netherlands.

DESIGN

Prospective cohort study.

METHODS

The right eye of healthy subjects was examined with the following 6 devices: IOLMaster (automated keratometry), Lenstar (automated keratometry), SMI Reference Unit 3 (automated keratometry), Javal (manual keratometry), KR-1W (corneal topography), and Pentacam (Scheimpflug imaging). An experienced operator obtained 3 repeated measurements. An inexperienced operator obtained additional measurements with the SMI Reference Unit 3. Astigmatism vector analysis was used to determine the comparability, repeatability, and interobserver variability.

RESULTS

Corneal astigmatism vectors measured by automated, manual, or simulated keratometry were comparable except for the Pentacam equivalent keratometry (K) (P<.001, repeated-measures analysis of variance [ANOVA]). The mean difference between the equivalent K and other K values was 0.18 to 0.29 diopter (D) (P<.05, Hotelling trace multivariate ANOVA). The mean differences between automated, manual, and simulated keratometry were small (≤0.12 D). The within-subject standard deviation ranged from 0.05 D @ 21 degrees (KR-1W) to 0.18 D @ 23 degrees (Lenstar). The SMI Reference Unit showed small mean differences and comparable repeatability between the experienced operator and the inexperienced operator.

CONCLUSIONS

Vector analysis showed comparable corneal astigmatism measurements using automated, manual, and simulated keratometry. Pentacam equivalent K values were not comparable with those of the other keratometers. The repeatability of astigmatism magnitudes was acceptable; however, the repeatability of astigmatism meridians was moderate. The SMI Reference Unit showed good interobserver variability.

摘要

目的

确定不同设备测量角膜散光的可比性和可重复性,并确定新型自动角膜曲率计的观察者间变异性。

设置

荷兰马斯特里赫特大学眼科诊所。

设计

前瞻性队列研究。

方法

对健康受试者的右眼进行以下 6 种设备的检查:IOLMaster(自动角膜曲率计)、Lenstar(自动角膜曲率计)、SMI Reference Unit 3(自动角膜曲率计)、Javal(手动角膜曲率计)、KR-1W(角膜地形图)和 Pentacam(Scheimpflug 成像)。一位有经验的操作人员进行了 3 次重复测量。一位缺乏经验的操作人员对 SMI Reference Unit 3 进行了额外的测量。使用散光矢量分析来确定可比性、重复性和观察者间变异性。

结果

除 Pentacam 等效角膜曲率计(K)外(P<.001,重复测量方差分析[ANOVA]),自动、手动或模拟角膜曲率计测量的角膜散光矢量是可比的(P<.001,重复测量方差分析[ANOVA])。等效 K 与其他 K 值之间的平均差值为 0.18 至 0.29 屈光度(D)(P<.05,Hotelling 迹多元方差分析)。自动、手动和模拟角膜曲率计之间的平均差异较小(≤0.12 D)。个体内标准差范围为 0.05 D @ 21 度(KR-1W)至 0.18 D @ 23 度(Lenstar)。SMI Reference Unit 显示出经验丰富的操作人员和缺乏经验的操作人员之间的小平均差异和可比重复性。

结论

向量分析显示,使用自动、手动和模拟角膜曲率计测量的角膜散光具有可比性。Pentacam 等效 K 值与其他角膜曲率计不可比。散光幅度的重复性可接受;然而,散光子午线的重复性中等。SMI Reference Unit 显示出良好的观察者间变异性。

文献检索

告别复杂PubMed语法,用中文像聊天一样搜索,搜遍4000万医学文献。AI智能推荐,让科研检索更轻松。

立即免费搜索

文件翻译

保留排版,准确专业,支持PDF/Word/PPT等文件格式,支持 12+语言互译。

免费翻译文档

深度研究

AI帮你快速写综述,25分钟生成高质量综述,智能提取关键信息,辅助科研写作。

立即免费体验