Pentacam 高分辨率角膜曲率计评估与自动角膜曲率计和传统角膜地形测量的比较。

Keratometry evaluations with the Pentacam high resolution in comparison with the automated keratometry and conventional corneal topography.

机构信息

Medical and Health Science Center, Department of Ophthalmology, University of Debrecen, Debrecen, Hungary.

出版信息

Cornea. 2012 Jan;31(1):36-41. doi: 10.1097/ICO.0b013e318204c666.

Abstract

PURPOSE

To determine the reliability and repeatability of keratometry (K) measurements obtained with the Pentacam high resolution (HR), automated keratometry, and corneal topography systems.

METHODS

The right eyes of 46 healthy subjects were examined prospectively. Keratometry measurements in the flat (Kf) and steep (Ks) meridians were taken by 2 independent investigators with the Pentacam HR (Oculus, Wetzlar, Germany) followed by automated keratorefractometry (KR-8100; Topcon, Tokyo, Japan), and corneal topography (TMS-4; Tomey, Erlangen, Germany).

RESULTS

The mean K readings of the Pentacam HR, automated keratometry, and corneal topography were 43.40/43.34 diopter (D), 43.99/43.98 D, and 43.80/43.83 D, respectively. The difference between the values was statistically significant (P < 0.0001, repeated measures analysis of variance). Strong significant correlation was observed between the Pentacam HR and keratometry (Kf: r = 0.952/0.954; Ks: r = 0.845, Spearman rank test), and Pentacam HR and corneal topography (Kf: r = 0.933/0.930; Ks: r = 0.838/0.829) (P < 0.0001). No significant difference was presented between the 2 investigators for any of the instruments (P = 0.215-0.983). Moreover, high correlation was found between the K readings of the observers (interoperator intraclass correlation coefficients ranged from 0.95 to 0.99).

CONCLUSIONS

The Pentacam HR provided reliable K measurements in clinical practice in comparison with an automated keratometer and a corneal topographer. Based on the results, for patient follow-up, one keratometry device is recommended.

摘要

目的

确定 Pentacam 高分辨率(HR)、自动角膜曲率计和角膜地形图系统获得的角膜曲率(K)测量的可靠性和可重复性。

方法

前瞻性检查 46 名健康受试者的右眼。由 2 名独立研究人员使用 Pentacam HR(德国 Oculus,Wetzlar)分别在平坦(Kf)和陡峭(Ks)子午线处进行角膜曲率测量,然后使用自动角膜屈光计(KR-8100;日本 Topcon)和角膜地形图(Tomey,德国 Erlangen)进行角膜曲率测量。

结果

Pentacam HR、自动角膜曲率计和角膜地形图的平均 K 值分别为 43.40/43.34 屈光度(D)、43.99/43.98 D 和 43.80/43.83 D。这些值之间的差异具有统计学意义(P < 0.0001,重复测量方差分析)。Pentacam HR 与角膜曲率计(Kf:r = 0.952/0.954;Ks:r = 0.845,Spearman 秩检验)和 Pentacam HR 与角膜地形图(Kf:r = 0.933/0.930;Ks:r = 0.838/0.829)之间存在显著的强相关性(P < 0.0001)。任何仪器的 2 位研究者之间均无显著差异(P = 0.215-0.983)。此外,观察者的 K 值之间存在高度相关性(观察者之间的组内相关系数范围为 0.95 至 0.99)。

结论

与自动角膜曲率计和角膜地形图相比,Pentacam HR 在临床实践中提供了可靠的 K 测量值。根据这些结果,建议在患者随访中使用一种角膜曲率计设备。