Suppr超能文献

基于测量的总角膜屈光力评估基于角膜曲率的 toric 人工晶状体的准确性。

An evaluation of the accuracy of toric intraocular lens power calculation based on measured total corneal refractive power.

机构信息

Department of Cataract, Shenyang Aier Eye Optometry Hospital, Shenyang, China.

出版信息

Indian J Ophthalmol. 2023 Feb;71(2):541-546. doi: 10.4103/ijo.IJO_1539_22.

Abstract

PURPOSE

To evaluate a method using measured values of total corneal refractive power (TCRP) for a manufacturer's online calculator by comparing it with the Barrett toric calculator (BTC) and Kane toric calculator (KTC) combined with simulated keratometry values (SimK).

METHODS

This was a retrospective case series. Patient records were reviewed to identify the patients who had biometry with the IOL Master 700 and Pentacam recorded before toric IOL implantation and refractive follow-up data after implantation. The predicted error in residual astigmatism was calculated by vector analysis according to the calculation methods and the measurements used.

RESULTS

A total of 70 eyes of 56 patients were included. The mean absolute astigmatism prediction errors were 0.6 ± 0.32, 0.59 ± 0.35, and 0.61 ± 0.35 D for the ATC, BTC, and KTC calculators, respectively (P = 0.934), and the centroid of the prediction errors were 0.3 D @ 178°, 0.11 D @ 102°, and 0.09 D @ 147°, respectively (P = 0.23). In the with-the-rule subgroup, the centroid of the prediction error was 0.34 D @ 176° for ATC and was the highest among the three calculation methods (P = 0.046).

CONCLUSION

The ATC, BTC, and KTC calculators had similar performance with regards to their astigmatism prediction accuracy. The ATC calculator combined with 4.0-mm apex/ring readings of TCRP was slightly intended to result in against-the-rule residual astigmatism.

摘要

目的

通过与 Barrett 散光计算器(BTC)和 Kane 散光计算器(KTC)结合模拟角膜曲率值(SimK)比较,评估制造商在线计算器中使用总角膜屈光力(TCRP)实测值的方法。

方法

这是一项回顾性病例系列研究。回顾患者记录,以确定在植入散光人工晶状体前用 IOLMaster700 和 Pentacam 进行生物测量以及植入后进行屈光随访的患者。根据使用的计算方法和测量值,通过向量分析计算残余散光的预测误差。

结果

共纳入 56 例 70 只眼。ATC、BTC 和 KTC 计算器的平均绝对散光预测误差分别为 0.6±0.32、0.59±0.35 和 0.61±0.35D(P=0.934),预测误差的质心分别为 0.3D@178°、0.11D@102°和 0.09D@147°(P=0.23)。在顺规亚组中,ATC 的预测误差质心为 0.34D@176°,在三种计算方法中最高(P=0.046)。

结论

ATC、BTC 和 KTC 计算器在散光预测准确性方面具有相似的性能。ATC 计算器结合 4.0mm 顶点/环 TCRP 读数,略微倾向于产生逆规残余散光。

https://cdn.ncbi.nlm.nih.gov/pmc/blobs/8bc3/10228908/e1e521b11931/IJO-71-541-g001.jpg

文献AI研究员

20分钟写一篇综述,助力文献阅读效率提升50倍。

立即体验

用中文搜PubMed

大模型驱动的PubMed中文搜索引擎

马上搜索

文档翻译

学术文献翻译模型,支持多种主流文档格式。

立即体验