An evaluation of the accuracy of toric intraocular lens power calculation based on measured total corneal refractive power.

机构信息

Department of Cataract, Shenyang Aier Eye Optometry Hospital, Shenyang, China.

出版信息

Indian J Ophthalmol. 2023 Feb;71(2):541-546. doi: 10.4103/ijo.IJO_1539_22.

Abstract

PURPOSE

To evaluate a method using measured values of total corneal refractive power (TCRP) for a manufacturer's online calculator by comparing it with the Barrett toric calculator (BTC) and Kane toric calculator (KTC) combined with simulated keratometry values (SimK).

METHODS

This was a retrospective case series. Patient records were reviewed to identify the patients who had biometry with the IOL Master 700 and Pentacam recorded before toric IOL implantation and refractive follow-up data after implantation. The predicted error in residual astigmatism was calculated by vector analysis according to the calculation methods and the measurements used.

RESULTS

A total of 70 eyes of 56 patients were included. The mean absolute astigmatism prediction errors were 0.6 ± 0.32, 0.59 ± 0.35, and 0.61 ± 0.35 D for the ATC, BTC, and KTC calculators, respectively (P = 0.934), and the centroid of the prediction errors were 0.3 D @ 178°, 0.11 D @ 102°, and 0.09 D @ 147°, respectively (P = 0.23). In the with-the-rule subgroup, the centroid of the prediction error was 0.34 D @ 176° for ATC and was the highest among the three calculation methods (P = 0.046).

CONCLUSION

The ATC, BTC, and KTC calculators had similar performance with regards to their astigmatism prediction accuracy. The ATC calculator combined with 4.0-mm apex/ring readings of TCRP was slightly intended to result in against-the-rule residual astigmatism.

摘要

目的

通过与 Barrett 散光计算器(BTC)和 Kane 散光计算器(KTC)结合模拟角膜曲率值(SimK)比较,评估制造商在线计算器中使用总角膜屈光力(TCRP)实测值的方法。

方法

这是一项回顾性病例系列研究。回顾患者记录,以确定在植入散光人工晶状体前用 IOLMaster700 和 Pentacam 进行生物测量以及植入后进行屈光随访的患者。根据使用的计算方法和测量值,通过向量分析计算残余散光的预测误差。

结果

共纳入 56 例 70 只眼。ATC、BTC 和 KTC 计算器的平均绝对散光预测误差分别为 0.6±0.32、0.59±0.35 和 0.61±0.35D(P=0.934),预测误差的质心分别为 0.3D@178°、0.11D@102°和 0.09D@147°(P=0.23)。在顺规亚组中,ATC 的预测误差质心为 0.34D@176°,在三种计算方法中最高(P=0.046)。

结论

ATC、BTC 和 KTC 计算器在散光预测准确性方面具有相似的性能。ATC 计算器结合 4.0mm 顶点/环 TCRP 读数,略微倾向于产生逆规残余散光。