Department of Pediatrics, Asahikawa Medical University, Asahikawa, Hokkaido, Japan.
J Clin Lab Anal. 2012 Sep;26(5):328-35. doi: 10.1002/jcla.21527.
Serum growth hormone (GH) level is measured largely through immunoassays in clinical practice. However, a few cases with bioinactive and immunoreactive GH have also been reported. We describe here a new bioassay system for GH determination using the BaF/GM cell line, which proliferates in a dose-dependent manner on hGH addition; cell proliferation was blocked by anti-hGH antibody. This bioassay had the lowest detection limit (∼0.02 ng/ml) reported thus far and the highest specificity for GH. The bioassay results were compared with those of an immunoradiometric assay across 163 patient samples in various endocrine states. A close correlation (the ratio of bioactivity/immunoreactivity was 1.04 ± 0.33, mean ± SD) was observed between bioactivity and immunoreactivity in these samples. The newly developed system is a specific, sensitive, easy, and fast bioassay system for GH determination; we consider it useful for evaluating GH bioactivity in various endocrine states.
血清生长激素 (GH) 水平在临床实践中主要通过免疫测定法来测量。然而,也有一些报道称存在生物活性和免疫反应性 GH。我们在这里描述了一种新的 GH 测定生物分析系统,该系统使用 BaF/GM 细胞系,该细胞系在添加 hGH 的情况下呈剂量依赖性增殖;细胞增殖被抗 hGH 抗体阻断。该生物分析具有迄今为止报道的最低检测限(约 0.02ng/ml)和对 GH 的最高特异性。将生物分析结果与 163 例各种内分泌状态患者样本的免疫放射测定结果进行比较。在这些样本中,生物活性和免疫反应性之间观察到密切相关(生物活性/免疫反应性的比值为 1.04±0.33,平均值±标准差)。新开发的系统是一种用于 GH 测定的特异性、灵敏性、简便、快速的生物分析系统;我们认为它在评估各种内分泌状态下的 GH 生物活性方面很有用。