Melton L A, Tracy M L, Möller G
California Veterinary Diagnostic Laboratory System, University of California, Davis 95617.
Clin Chem. 1990 Feb;36(2):247-50.
This rapid, accurate procedure for trace elements and electrolytes in serum requires little sample preparation: to 1 mL of serum a single reagent is added that contains trichloroacetic and hydrochloric acids for protein precipitation, hydroxylamine sulfate for iron reduction, and yttrium as the internal standard. After centrifugation, the supernates are directly analyzed for Na, K, Mg, Ca, Pi, Fe, Cu, and Zn by inductively-coupled plasma emission spectrometry. The CVs were respectively 7.9%, 8.4%, 8.6%, 10.0%, 9.0%, 9.4%, 9.0%, and 9.0% for five assays of National Institute of Standards and Technology Standard Reference Material (SRM) no. 1598, Bovine Serum. Analytical recoveries ranged from 92% to 107% for both SRM 1598 and commercial control serum.
这种用于血清中微量元素和电解质的快速、准确的检测方法几乎不需要样品制备:向1 mL血清中加入一种单一试剂,该试剂含有用于蛋白质沉淀的三氯乙酸和盐酸、用于铁还原的硫酸羟胺以及作为内标物的钇。离心后,上清液直接通过电感耦合等离子体发射光谱法分析钠、钾、镁、钙、磷、铁、铜和锌。对于美国国家标准与技术研究院标准参考物质(SRM)1598号牛血清进行的五次检测,变异系数(CV)分别为7.9%、8.4%、8.6%、10.0%、9.0%、9.4%、9.0%和9.0%。SRM 1598和商业对照血清的分析回收率在92%至107%之间。