Aser Raed, Orhan Coskun, Niemann Bernd, Roth Peter, Perepelitsa Andre, Attmann Tim, Böning Andreas
Department of Cardiovascular Surgery, University Hospital Giessen, Giessen, Germany.
Thorac Cardiovasc Surg. 2014 Feb;62(1):66-9. doi: 10.1055/s-0032-1311544. Epub 2012 Oct 3.
To determine the pacing and sensing properties of different temporary epicardial pacemaker electrodes after cardiac surgery depending on position at the heart and time after surgery.
From September 2009 to October 2010, 60 patients undergoing cardiac surgery were prospectively randomized into two groups: group O: Osypka-electrodes (n = 30), group M: Medtronic-electrodes (n = 30). In position 1, the bipolar electrodes were inserted onto the anterior wall of the right ventricle and at the right atrial auricle, in position 2, onto the diaphragmal wall of the right ventricle and at the aortic aspect of the superior vena cava medial close to the atrium. Sensing values and pacing thresholds were measured for all electrodes during surgery, on day 1 and every second day up to day 10 after surgery.
In both groups, pacing thresholds (both positions) were higher during surgery (ventricle 3.1 ± 0.6 V, atrium 3.1 ± 0.3 V) than at day 1 (ventricle 2.4 ± 0.7 V, atrium 2.4 ± 0.3 V) and increased during the perioperative course until day 10 (ventricle 4.7 ± 1.0 V, atrium 4.9 ± 1.1 V, p = 0.04, p = 0.02). P and R wave amplitudes did not change over time (atrium 5.1 ± 0.1 mV initially, 4.2 ± 0.1 mV at removal (p = ns); ventricle 10.4 ± 0.2 mV vs. 10.1 ± 0.25 mV). Group M had better median pacing thresholds compared with group O (atrium: 2.9 ± 0.6 V vs. 3.9 ± 0.7 V, p = 0.04 and ventricle: 2.6 ± 0.6 V vs. 3.9 ± 0.6 V, p = 0.045). Atrial position 1 was superior to position 2 concerning pacing thresholds of Medtronic electrodes (2.1 ± 0.3 mV vs. 3.4 ± 0.4 mV, p = 0.02). Osypka-electrodes were easier to handle due to their more pliable texture.
根据心脏手术后不同临时心外膜起搏电极在心脏的位置及术后时间,确定其起搏和感知特性。
2009年9月至2010年10月,60例行心脏手术的患者被前瞻性随机分为两组:O组:奥西普卡电极(n = 30),M组:美敦力电极(n = 30)。在位置1,将双极电极插入右心室前壁和右心耳;在位置2,插入右心室膈面壁以及靠近心房处上腔静脉内侧的主动脉侧。在手术期间、术后第1天以及术后第10天之前每隔一天测量所有电极的感知值和起搏阈值。
两组中,手术期间(心室3.1±0.6V,心房3.1±0.3V)的起搏阈值均高于术后第1天(心室2.4±0.7V,心房2.4±0.3V),并在围手术期过程中直至第10天有所升高(心室4.7±1.0V,心房4.9±1.1V,p = 0.04,p = 0.02)。P波和R波振幅随时间无变化(心房最初5.1±0.1mV,拔除时4.2±0.1mV(p = 无统计学意义);心室10.4±0.2mV对10.1±0.25mV)。与O组相比,M组的起搏阈值中位数更好(心房:2.9±0.6V对3.9±0.7V,p = 0.04;心室:2.6±0.6V对3.9±0.6V,p = 0.045)。就美敦力电极的起搏阈值而言,心房位置1优于位置2(2.1±0.3mV对3.4±0.4mV,p = 0.02)。奥西普卡电极质地更柔韧,更易于操作。