Med Biol Eng Comput. 2012 Dec;50(12):1239-47. doi: 10.1007/s11517-012-0955-z. Epub 2012 Sep 30.
Measurement of flow-mediated dilatation (FMD) is the conventional non-invasive method for assessment of endothelial function; however, it requires an expensive ultrasound system and high levels of technical skill. Therefore, we developed a novel method for measurement of endothelial function, namely, measurement of ezFMD. ezFMD estimates the degree of vasodilatation from the oscillation signals transmitted to a sphygmomanometer cuff attached to the upper arm. The objective of this study was to validate the principle underlying the measurement of ezFMD, and to evaluate the repeatability of the ezFMD measurements. We observed the blood vessel behavior and oscillometric pattern in ten subjects. When the cuff was inflated to the level of the mean blood pressure, the oscillation amplitude increased with increasing degree of vasodilatation. In experiment to evaluate the repeatability of the ezFMD measurement, the average difference between the paired measurements was 3.7 %, the standard deviation was 11.5 %, and the average coefficient of variation value for the 11 paired measurements was 23.7 %. These results suggest the validity of the principle underlying the measurement of the ezFMD for the assessment of endothelial function. And, this study suggests that the repeatability of the ezFMD measurements is superior to that of the conventional measurement of FMD.
血流介导的扩张(FMD)的测量是评估内皮功能的传统非侵入性方法;然而,它需要昂贵的超声系统和高水平的技术技能。因此,我们开发了一种新的内皮功能测量方法,即 ezFMD 的测量。ezFMD 从附加在上臂的血压袖带传输的振荡信号来估计血管扩张的程度。本研究的目的是验证 ezFMD 测量原理,并评估 ezFMD 测量的可重复性。我们观察了 10 名受试者的血管行为和示波模式。当袖带充气到平均血压水平时,随着血管扩张程度的增加,振荡幅度增加。在评估 ezFMD 测量重复性的实验中,配对测量之间的平均差异为 3.7%,标准差为 11.5%,11 对配对测量的平均变异系数值为 23.7%。这些结果表明,ezFMD 测量评估内皮功能的原理是有效的。并且,这项研究表明,ezFMD 测量的重复性优于传统的 FMD 测量。