应用新型角膜地形图仪评估干眼的泪膜稳定性。

Assessment of tear film stability in dry eye with a newly developed keratograph.

机构信息

Department of Ophthalmology and Visual Science, Eye, Ear, Nose, and Throat Hospital, School of Shanghai Medicine, Fudan University, Shanghai, China.

出版信息

Cornea. 2013 May;32(5):716-21. doi: 10.1097/ICO.0b013e3182714425.

Abstract

PURPOSE

To investigate the applicability of a newly developed corneal topographer in assessing tear film stability.

METHODS

This is a prospective, case-control study. Forty-four Chinese dry eye patients and 41 normal subjects were recruited. Noninvasive tear film break-up time (NI-BUT) was measured using a new method based on a corneal topographer equipped with modified scan software. The reliability of the measurements was determined. Then, the correlations between the NI-BUT and the traditional fluorescein tear film break-up time, Schirmer I test values, and inferior tear meniscus height measurements were determined. The receiver operating characteristic curve technique was used to evaluate the NI-BUT examination in the diagnosis of dry eye.

RESULTS

In total, a significant difference between the NI-BUT and the fluorescein tear film break-up time was found (3.2 ± 2.3 seconds vs. 5.2 ± 3.4 seconds; P < 0.001). The coefficient of variation and intraclass correlation coefficient values of NI-BUT were 12.8% and 0.93, respectively, for NI-BUT for intraobserver repeatability and 15.4% and 0.88, respectively for interobserver repeatability. The NI-BUT showed a good correlation with other dry eye examinations (all P < 0.05). In addition, the NI-BUT was significantly shorter in dry eye patients (2.0 ± 0.2 seconds) than in normal subjects (4.3 ± 0.3 seconds; P < 0.001). When the cutoff value was set at <2.65 seconds, the sensitivity and specificity of the test were 84.1% and 75.6%, respectively.

CONCLUSIONS

Measurements of NI-BUT obtained with the newly developed corneal topographer may provide a simple, noninvasive screening test for dry eyes with acceptable sensitivity, specificity, and repeatability.

摘要

目的

研究一种新开发的角膜地形图仪在评估泪膜稳定性方面的适用性。

方法

这是一项前瞻性病例对照研究。共招募了 44 例中国干眼患者和 41 例正常对照者。使用一种新方法,基于配备改良扫描软件的角膜地形图仪,对非侵入性泪膜破裂时间(NI-BUT)进行测量。确定了测量的可靠性。然后,确定了 NI-BUT 与传统荧光素泪膜破裂时间、Schirmer I 试验值和下泪膜高度测量值之间的相关性。使用受试者工作特征曲线技术评估 NI-BUT 检查在干眼诊断中的作用。

结果

NI-BUT 与荧光素泪膜破裂时间之间存在显著差异(3.2 ± 2.3 秒 vs. 5.2 ± 3.4 秒;P < 0.001)。NI-BUT 的变异系数和组内相关系数分别为 12.8%和 0.93,用于观察者内重复性;分别为 15.4%和 0.88,用于观察者间重复性。NI-BUT 与其他干眼检查具有良好的相关性(均 P < 0.05)。此外,干眼患者的 NI-BUT 明显短于正常对照者(2.0 ± 0.2 秒 vs. 4.3 ± 0.3 秒;P < 0.001)。当截断值设定为 <2.65 秒时,该检测的敏感性和特异性分别为 84.1%和 75.6%。

结论

使用新开发的角膜地形图仪测量 NI-BUT 可能为干眼提供一种简单、非侵入性的筛查试验,具有可接受的敏感性、特异性和重复性。