杜罗姆髋臼组件的变形及其在尸体模型中对摩擦学的影响——一项模拟器研究

Deformation of the Durom acetabular component and its impact on tribology in a cadaveric model--a simulator study.

作者信息

Liu Feng, Chen Zhefeng, Gu Yanqing, Wang Qing, Cui Weiding, Fan Weimin

机构信息

Department of Orthopedics, the First Affiliated Hospital with Nanjing Medical University, Nanjing, China.

出版信息

PLoS One. 2012;7(10):e45786. doi: 10.1371/journal.pone.0045786. Epub 2012 Oct 29.

Abstract

BACKGROUND

Recent studies have shown that the acetabular component frequently becomes deformed during press-fit insertion. The aim of this study was to explore the deformation of the Durom cup after implantation and to clarify the impact of deformation on wear and ion release of the Durom large head metal-on-metal (MOM) total hips in simulators.

METHODS

Six Durom cups impacted into reamed acetabula of fresh cadavers were used as the experimental group and another 6 size-paired intact Durom cups constituted the control group. All 12 Durom MOM total hips were put through a 3 million cycle (MC) wear test in simulators.

RESULTS

The 6 cups in the experimental group were all deformed, with a mean deformation of 41.78 ± 8.86 µm. The average volumetric wear rate in the experimental group and in the control group in the first million cycle was 6.65 ± 0.29 mm(3)/MC and 0.89 ± 0.04 mm(3)/MC (t = 48.43, p = 0.000). The ion levels of Cr and Co in the experimental group were also higher than those in the control group before 2.0 MC. However there was no difference in the ion levels between 2.0 and 3.0 MC.

CONCLUSIONS

This finding implies that the non-modular acetabular component of Durom total hip prosthesis is likely to become deformed during press-fit insertion, and that the deformation will result in increased volumetric wear and increased ion release.

CLINICAL RELEVANCE

This study was determined to explore the deformation of the Durom cup after implantation and to clarify the impact of deformation on wear and ion release of the prosthesis. Deformation of the cup after implantation increases the wear of MOM bearings and the resulting ion levels. The clinical use of the Durom large head prosthesis should be with great care.

摘要

背景

近期研究表明,压配植入过程中髋臼部件常发生变形。本研究旨在探讨Durom髋臼杯植入后的变形情况,并阐明变形对Durom大头金属对金属(MOM)全髋关节在模拟器中磨损及离子释放的影响。

方法

将六个压配入新鲜尸体扩髓髋臼的Durom髋臼杯作为实验组,另外六个尺寸配对的完整Durom髋臼杯作为对照组。所有12个Durom MOM全髋关节在模拟器中进行300万次循环(MC)的磨损试验。

结果

实验组的6个髋臼杯均发生变形,平均变形量为41.78±8.86µm。实验组和对照组在第一个100万次循环中的平均体积磨损率分别为6.65±0.29mm³/MC和0.89±0.04mm³/MC(t = 48.43,p = 0.000)。在2.0MC之前,实验组中Cr和Co的离子水平也高于对照组。然而,在2.0MC和3.0MC之间离子水平没有差异。

结论

这一发现表明,Durom全髋关节假体的非模块化髋臼部件在压配植入过程中可能会变形,并且这种变形会导致体积磨损增加和离子释放增加。

临床意义

本研究旨在探讨Durom髋臼杯植入后的变形情况,并阐明变形对假体磨损及离子释放的影响。植入后髋臼杯的变形会增加MOM轴承的磨损及由此产生的离子水平。Durom大头假体的临床应用应格外谨慎。